 National Instruments Corporation

3

NI USB-TC01 Calibration Procedure

Keep relative humidity below 80%.

Allow a warm up time of at least 15 minutes for the NI USB-TC01 and 
30 minutes for the calibrator

 

to ensure that the measurement circuitry 

is at a stable operating temperature.

Calibration Procedures

The calibration process includes the following steps:

1.

Initial Setup

Install the device.

2.

Verification

—Verify the existing operation of the device. This step 

confirms whether the device is operating within the published 
specifications prior to adjustment.

3.

Adjustment

—If the device does not fall within the desired 

specifications, you can return the device to NI for a factory test to 
adjust the calibration constants. Go to 

ni.com/calibration

 for 

information about returning a device to NI for a factory test.

4.

Reverification

—Repeat the verification procedure to ensure that the 

device is operating within the published specifications after adjustment.

The first two procedures are described in more detail in the following 
sections.

Initial Setup

Refer to the 

Getting Started with the NI USB-TC01

 for information about 

how to install the hardware.

Verification

This section provides instructions for verifying the NI USB-TC01

 

specifications. The limits are based on the specifications that the device 
should meet if it has been one year since the original factory calibration.

Verifying Temperature Measurement Accuracy

Complete the following steps to verify temperature measurement accuracy.

1.

Connect the thermocouple output of the calibrator to the NI USB-TC01.

2.

Set the calibrator to output a 0 °C type J measurement.

3.

Acquire a temperature measurement with the NI USB-TC01.

4.

Verify that the NI USB-TC01 measurement is 0 ±2.05 °C.

5.

Set the calibrator to output a 1000 °C type J measurement.

6.

Acquire a temperature measurement with the NI USB-TC01.

7.

Verify that the NI USB-TC01 measurement is 1000 ±1.96 °C.