NI USB-TC01 Calibration Procedure

2

ni.com

Documentation Requirements

For information about the NI USB-TC01, you can consult the following 
documents: 

Getting Started with the NI USB-TC01

—provides instructions for 

installing and configuring the NI USB-TC01.

NI USB-TC01 User Guide and Specifications

—includes detailed 

information about NI USB-TC01 and provides the published 
specification values for the NI USB-TC01. Refer to the most recent 

NI USB-TC01 User Guide and Specifications 

online at 

ni.com/

manuals

 for specifications.

You can also find the latest versions of the documentation at 

ni.com/

manuals

.

Test Equipment

National Instruments recommends that you use the equipment in Table 
for calibrating the NI USB-TC01.

Test Conditions

The following setup and environmental conditions are required to ensure 
the NI USB-TC01

 

meets published specifications. 

Keep connections to the device as short as possible. Long cables and wires 
act as antennae, picking up extra noise that can affect measurements.

Verify that all connections to the device are secure.

Maintain an ambient temperature of 25 ±5 °C.

Table 1.  

Recommended Equipment

Equipment

Recommended Model

Requirements

Calibrator

Fluke 5500A

If this instrument is unavailable, use a 
high-precision simulated thermocouple 
output module capable of type J with an 
accuracy of ±0.4 °C at 0 °C and 1000 °C 
simulated outputs.

Thermocouple wire

Type J

Two miniature 
thermocouple 
connectors

Type J