SCXI-1102/B/C Calibration Procedure
4.
Set the SCXI-1102/B/C to a gain value of 1 or 100 using the DAQmx
Setup SCXI Calibration VI.
5.
Set the calibrator to the desired voltage for the gain value you set in the
previous step. Refer to Table 3 or 4 for appropriate voltage set points.
6.
Wait for the settling time to elapse and then record the measured output
from the DMM. Refer to Table 5 for the settling time of the
SCXI-1102/B/C.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup1102Cal
with the following
parameters:
calHandle
:
calHandle
channelNames
:
"SC1Mod1/ai
X
"
†
gain
:
1
or
100
†
Where
X
refers to the channel you are calibrating.
Table 3.
SCXI-1102 Voltage Set Points
Gain
Set Points (V)
1
–9
–6
–2
0
2
6
9
100
–0.09
–0.06
–0.02
0.00
0.02
0.06
0.09
Table 4.
SCXI-1102 B and SCXI-1102 C Voltage Set Points
Gain
Set Points (V)
1
–9
–6
–3
0
3
6
9
100
–0.09
–0.06
–0.03
0.00
0.03
0.06
0.09
Table 5.
Settling Time of the SCXI-1102/B/C
SCXI-1102
SCXI-1102B, SCXI-1102C
60 seconds
5 second