National Instruments Corporation

3

SCXI-1338 Installation Guide

Do

not

operate damaged equipment. The safety protection features built into this device

can become impaired if the device becomes damaged in any way. If the device is damaged,
turn the device off and do

not

use it until service-trained personnel can check its safety.

If necessary, return the device to National Instruments for service and repair to ensure that
its safety is not compromised.

Do

not

operate this equipment in a manner that contradicts the information specified in this

document. Misuse of this equipment could result in a shock hazard.

Do

not

substitute parts or modify equipment. Because of the danger of introducing

additional hazards, do

not

install unauthorized parts or modify the device. Return the

device to National Instruments for service and repair to ensure that its safety features are
not compromised.

You

must

insulate all of your signal connections to the highest voltage with which the

SCXI-1338 can come in contact.

When using the device with high common-mode voltages, you

must

insulate your signal

wires for the highest input voltage. National Instruments is

not

liable for any damages or

injuries resulting from inadequate signal wire insulation. Use only 26-14 AWG wire with
a minimum voltage rating of 300 V and a temperature value of 60 °C for measuring up
to 300 V.

When connecting or disconnecting signal lines to the SCXI terminal block screw terminals,
make sure the lines are powered off. Potential differences between the lines and the SCXI
ground can create a shock hazard while you connect the lines.

Connections, including power signals to ground and vice versa, that exceed any of the
maximum signal ratings on the SCXI device can create a shock or fire hazard, or can
damage any or all of the boards connected to the SCXI chassis, the host computer, and the
SCXI device. National Instruments is

not

liable for any damages or injuries resulting from

incorrect signal connections.

If hazardous voltages (

30 V

rms

and 42.4 V

peak

or 60 VDC) are present, you

must

connect

a safety earth-ground wire to the terminal block safety-ground lug, shown in Figure 2. This
complies with safety agency requirements and protects against electric shock when the
terminal block is not connected to the chassis. To connect the safety earth-ground to the
safety-ground lug, run an earth-ground wire from the signal source to the terminal block.
National Instruments is

not

liable for any damages or injuries resulting from inadequate

safety earth-ground connections.

Do

not

loosen or re-orient the safety-ground lug hardware when connecting the

safety-ground wire; to do so reduces the safety isolation between the high voltage and
safety ground.

Clean the module and accessories by brushing off light dust with a soft non-metallic brush.

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for SCXI-1338

Page 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Page 2: ...onnect to the SCXI 1120 SCXI 1120D SCXI 1125 and SCXI 1126 modules The SCXI 1338 has eight 249 Ω precision resistors to convert current to voltage when measuring 0 20 mA and 4 20 mA process current sources The terminal block has 18 screw terminals for easy connection Eight pairs of screw terminals are for signal connection to the eight module inputs One pair of terminals connects to the SCXI chass...

Page 3: ...s programs subprograms subroutines device names functions operations variables filenames and extensions and code excerpts Safety Information Caution Do not operate the device in an explosive atmosphere or where there may be flammable gases or fumes Keep away from live circuits Do not remove equipment covers or shields unless you are trained to do so If signal wires are connected to the device haza...

Page 4: ...up to 300 V When connecting or disconnecting signal lines to the SCXI terminal block screw terminals make sure the lines are powered off Potential differences between the lines and the SCXI ground can create a shock hazard while you connect the lines Connections including power signals to ground and vice versa that exceed any of the maximum signal ratings on the SCXI device can create a shock or f...

Page 5: ...cquires the voltage To convert the voltage read by the SCXI module to the actual current being measured use the following formula current measured voltage 249 The voltage measured for a 4 20 mA current source is in the range of 0 996 4 98 VDC The SCXI 1338 can measure both current sources and voltage sources When measuring voltage signals you must remove the current loop resistor for that input ch...

Page 6: ...the signal wires through the strain relief opening You can add insulation or padding if necessary 4 Prepare your signal wire by stripping the insulation no more than 7 mm 5 Connect the wires to the screw terminals by inserting the stripped end of the wire fully into the terminal No bare wire should extend past the screw terminal Exposed wire increases the risk of a short circuit and equipment fail...

Page 7: ...ct the module front connector to its mating connector on the terminal block 2 Tighten the top and bottom thumbscrews on the back of the terminal block to hold it securely in place 1 Strain Relief Bar 2 Strain Relief Screws 3 Safety Ground Lug 4 Mating Connector 5 Thumbscrew 6 Top Cover Screws 7 Top Cover 3 4 2 1 5 6 7 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www ar...

Page 8: ...voltage signal common mode Channel to ground Each channel must remain within 300 Vrms or 300 VDC of ground Channel to channel Each channel must remain within 300 Vrms or 300 VDC of the voltage applied to any other channel Maximum input current 20 mA Field wiring connectivity Screw terminals for signals 16 8 pairs Screw terminals for functional earth ground 2 Safety earth ground lugs 1 Strain relie...

Page 9: ...elative humidity 10 to 90 noncondensing Pollution degree 2 Altitude maximum 2000 m Safety Designed in accordance with IEC 61010 1 UL 3111 1 and CAN CSA C22 2 No 1010 1 for electrical measuring and test equipment Installation Category II Pollution degree 2 Electromagnetic Compatibility EMC EMI CE C Tick and FCC Part 15 Class A Compliant Electrical emissions EN 55011 Class A at 10 m FCC Part 15A abo...

Page 10: ...port phone numbers e mail addresses and current events If you have searched the technical support resources on our Web site and still cannot find the answers you need contact your local office or National Instruments corporate For telephone support in the United States dial 512 795 8248 For telephone support outside the United States contact your local branch office Australia 03 9879 5166 Austria ...

Page 11: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Reviews: