NI SCXI-1332 Installation Instructions
Specifications
Maximum Voltage
Channel-to-ground ................................. 300 V, CAT II
Caution
This module is rated for Measurement Category II and intended to carry signal
voltages no greater than 300 V. This module features 1390 V
rms
isolation between input
signals and the chassis as verified by a dielectric withstand test, 1 minute maximum. Do
not
use this module for connection to signals or for measurements within Categories III or IV.
Refer to the
Read Me First: Safety and Radio-Frequency Interference
document for more
information about measurement categories.
Maximum Current
Maximum current
(per channel) .......................................... 2 ADC, 2 AAC
Environment
The NI SCXI-1332 is intended for indoor use only.
Operating temperature............................ 0 °C to 50 °C
Storage temperature ............................... –20 °C to 70 °C
Relative humidity ................................... 10% to 90% RH, noncondensing
Pollution Degree .................................... 2
Approved at altitudes up to 2,000 m.
Safety
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1
•
CAN/CSA-C22.2 No. 61010-1
Note
For UL and other safety certifications, refer to the product label or visit
ni.com/certification
, search by model or product line, and click the appropriate link
in the Certification column.