SCXI-1120/D Calibration Procedure
14
ni.com
7.
Adjust the input potentiometer of channel 0 until the DMM reading is
0 ±6.0 mV. Refer to Figure 3 for the potentiometer location and Table 6
for the potentiometer reference designator.
8.
Repeat steps 1 through 7 for the remaining analog inputs.
You have completed adjusting the module.
Verifying Adjusted Values
After you complete the adjustment procedure, it is important to verify the
accuracy of the adjusted values by repeating the procedure in the
section. Verifying the adjusted values ensures
the module is operating within its specifications after adjustments.
Note
If the SCXI-1120/D module fails after calibration, return it to NI for repair or
replacement.