PXIe-4330/4331 Calibration Procedure
Connect the Calibration and Switch Cables
Figure 5.
Connecting the Calibration Cable
1.
Screw the two standoffs into the PXIe-4330/4331 module and tighten the standoffs to
1.30 N · m (11.5 lb · in.), as shown in Figure 5.
2.
Connect the male end of the calibration cable to the CAL-4330 and tighten the two jack
screws to 0.90 N · m (8.0 lb · in.), as shown in Figure 5.
3.
Connect the female end of the calibration cable to the PXIe-4330/4331 and tighten the two
jack screws to 0.90 N · m (8.0 lb · in.), as shown in Figure 5.
1
CAL-4330
2
PXI Chassis
3
PXIe-4330/4331 (DUT)
4
Standoffs (2)
5
Jack Screws (2)
6
CAL-4330 to PXIe-4330/4331 Calibration Cable
7
Jack Screws (2)