PXIe-4330/4331 Calibration Procedure
|
© National Instruments
|
19
Table 8.
Configuration Settings for Full-Bridge Operation
Property
Value
Channel Name
Use channel names specific to your application.
Acquisition Mode
Finite number of samples
Rate (Hz)
25,600 or 102,400
*
Samples to Read
10,000 or 40,000
†
Measurement Type
Bridge
Max Input Limit
0.025 or 0.10
**
Min Input Limit
-0.025 or -0.10
**
Units
V/V
Bridge Type
Full Bridge
Nominal Bridge Resistance
350
Ω
Vex Source
Internal
Vex Value
0.625, 1.0, 1.5, 2.0, 2.5, 2.75, 3.3, 5.0, 7.5, or 10.0
*
*
Use the excitation value and rate that corresponds to the configuration being verified.
†
Read 10,000 samples when verifying at the 25,600 Hz rate. Read 40,000 samples when verifying at the
102,400 Hz rate.
**
Use -0.025 and 0.025 for excitation >2.5 V. Use -0.10 and 0.10 for excitation
≤
2.5 V.
Table 9.
Configuration Settings for Shunt Calibration Resistance
Accuracy Verification
Property
Value
Channel Name
Use channel names specific to your application.
Acquisition Mode
Finite number of samples
Rate (Hz)
25,600
Samples to Read
10,000
Measurement Type
Bridge
Max Input Limit
0.10
Min Input Limit
-0.10
Units
V/V
Bridge Type
Full Bridge
Nominal Bridge Resistance
350
Ω