Connecting Linear Voltage Input Signals
Figure 8. Linear mV Voltage Connection
24 VDC
TC+ (0:3)
TC– (0:3)
REM-11120
+
–
x4
a1
a2
b1
b2
Shield
100 mV
+
–
Figure 9. Linear VDC Voltage Connection
24 VDC
AI+
AI–
5 V Source
REM-11120
+
–
x4
+
–
a1
a2
b1
b2
•
Connect sensors to each channel which supply a linear voltage in ranges of ±100 mV or
±5 VDC.
•
For mV sensors, cable lengths of up to 250 m (820 ft.) are possible.
•
For VDC sensors, use shielded twisted sensor cables no longer than 5 m (16 ft.).