![National Instruments PXIe-8240 User Manual Download Page 8](http://html1.mh-extra.com/html/national-instruments/pxie-8240/pxie-8240_user-manual_709086008.webp)
Caution
Clean the hardware with a soft, nonmetallic brush. Make sure the
hardware is completely dry and free from contaminants before returning it to
service.
Environmental
Operating Environment
Ambient temperature range
0 °C to 55 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2.)
Relative humidity range
10% to 90%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Maximum altitude
2,000 m (800 mbar)
Pollution Degree
2
Indoor use only.
Storage Environment
Ambient temperature range
-40 °C to 71 °C (Tested in accordance with
IEC-60068-2-1 and IEC-60068-2-2.)
Relative humidity range
5% to 95%, noncondensing (Tested in
accordance with IEC-60068-2-56.)
Shock and Vibration
Operating shock
30 g peak, half-sine, 11 ms pulse (Tested in
accordance with IEC-60068-2-27. Meets MIL-
PRF-28800F Class 2 limits.)
Random vibration
Operating
5 Hz to 500 Hz, 0.3 g
rms
Nonoperating
5 Hz to 500 Hz, 2.4 g
rms
(Tested in accordance
with IEC-60068-2-64. Nonoperating test
profile exceeds the requirements of MIL-
PRF-28800F, Class 3.)
PXIe-8240 User Manual
|
© National Instruments
|
7