Note
Ensure that any connected external hardware is in an idle state and
not transmitting during self-calibration. It is not recommended to perform
self-calibration while connected in direct loopback configurations.
Note
For the best image rejection and LO leakage performance, it is
recommended that you run Self Cal or Self Cal Range for only the image
rejection and LO leakage portion if the temperature changes beyond ±1 °C.
Performing a Self-Calibration Using NI-RFSA
NI recommends you perform the self-calibration using the NI-RFSA example when
you self-calibrate the module.
1. Select
Start
»
All Programs
»
National Instruments
»
LabVIEW
to launch
LabVIEW.
2. Launch the Example Finder and navigate to the example.
1. Select
Hardware Input and Output
»
Modular Instruments
»
NI-
RFSA
»
Utilities
»
RFSA Self Calibration
.
2. Open the example.
3. Complete the following steps to configure the example:
1. Select the device identifier assigned to the PXIe-3623 in MAX in the
[Resource Name]
drop-down menu.
2. Set Clock Source to
OnboardClock
.
3. Set Self Calibration Step Operations to
Perform All Self Calibration
Steps
.
4. Run the VI.
Software Locations
ni.com
20
PXIe-5832 Getting Started