splitter loss
=
amplitude level
-
splitter output 2 power
24. Disconnect the 50 Ω SMA terminator (m) from
splitter output 2
. Refer to the remaining
assembly as the
power sensor assembly
.
The
power sensor assembly
is used in the following procedures:
•
Verifying 50 Ω Bandwidth
•
Verifying 1 MΩ Bandwidth
Verification
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.
Verification of the PXIe-5172 is complete only after you have successfully completed all tests
in this section using the
As-Found Limits
.
Refer to the following figure for the names and locations of the PXIe-5172 front panel
connectors. You can find information about the functions of these connectors in the device
getting started guide.
12
|
ni.com
|
PXIe-5172 Calibration Procedure