![background image](http://html1.mh-extra.com/html/national-instruments/pxie-4610/pxie-4610_getting-started-manual_3390302006.webp)
6
|
ni.com
|
NI PXIe-4610 Power Amplifier Getting Started Guide
Figure 4 shows the connection of a floating signal source to an NI PXIe-4610 input.
Figure 4.
Connection for Floating Signal Source
Connect a Load to the NI PXIe-4610 Output
The NI PXIe-4610 outputs are differential and ground-referenced.
You can connect floating loads and balanced, ground-referenced loads to the NI PXIe-4610
outputs. Refer to Figure 8 for the NI PXIe-4610 front panel.
Note
Refer to the
NI PXIe-4610 Specifications
for information about the operating
output range, short-circuit, and overvoltage protection.
Figure 5 shows the connection of a floating load to an NI PXIe-4610 output.
Figure 5.
Connection for Floating Load
IN CH
x
S
ign
a
l
S
o
u
rce
NI PXIe-4610
+
–
+
–
OUT CH
x
+
OUT CH
x
–
+
–
NI PXIe-4610
+
–
8
Ω