NI PXIe-4480/4481 Calibration Procedure
|
© National Instruments
|
33
Reverification
After completing the adjustment procedure, repeat the
section to determine the
As Left status of the device. Some, but not all, performance characteristics are affected by
adjustment and would be subject to reverification after adjustment. Characteristics affected by
adjustment have As Left test limits in addition to As Found test limits. Characteristics unaffected
by adjustment solely have test limits with no additional qualification.
Note
If any test fails reverification after performing an adjustment, verify that you
have met the
before returning your device to NI. Refer to
for assistance in returning the device to NI.
Updating the Calibration Date without Performing
Adjustment
You can update the calibration date without performing the adjustment procedure. This might
be appropriate if the PXIe-4480/4481 successfully passed each of the verification procedures
within the As Left test limits
without adjustment
. Complete the following steps to update the
calibration date:
1.
Open a calibration session using the DAQmx Initialize External Calibration VI with the
following parameters:
•
deviceName
:
Dev_name
•
password
:
NI
•
calHandle
:
&myCalHandle
2.
Close the session with the DAQmx Close External Calibration VI with the following
parameters:
•
calHandle
:
myCalHandle
•
action
:
DAQmx_Val_Action_Commit
Specifications
Refer to the
PXIe-4480/4481 Specifications
for calibration intervals and detailed specification
information.