32
|
ni.com
|
NI PXIe-4480//4481 Calibration Procedure
11. Call
DAQmxAdjustDSAAICalWithGainAndCoupling
(DAQmx Adjust DSA AI
Calibration With Gain and Coupling VI from the DAQmx Adjust DSA AI Calibration
Polymorphic VI), using the following parameters:
calHandle
:
myCalHandle
coupling
:
DAQmx_Val_AC
gain:
0
referenceVoltage:
6.3
Note
Set the
gain
parameter to the value of the device gain setting being calibrated,
where 0, for example, represents 0 dB.
12. Configure the calibrator to output a 3.15 V
rms
signal at 1 kHz.
13. Repeat step 11 for the 6 dB gain range, substituting the
gain
parameter with 6 and the
referenceVoltage
parameter with 3.15.
14. Configure the calibrator to output a 630 mV
rms
signal at 1 kHz.
15. Repeat step 11 for the 20 dB gain range, substituting the
gain
parameter with 20 and the
referenceVoltage
parameter with 0.63.
16. Configure the calibrator to output a 315 mV
rms
signal at 1 kHz.
17. Repeat step 11 for the 26 dB gain range, substituting the
gain
parameter with 26 and the
referenceVoltage
parameter with 0.315.
Note
All ranges must be adjusted for gain for the calibration to be valid.
18. Repeat steps 8 through 17 for each input channel, changing the cable connections
accordingly and substituting the
physical channels
parameter with the channel being
calibrated.
Note
All channels must be adjusted for gain for the calibration to be valid.
Committal
19. Call
DAQmxCloseExtCal
(DAQmx Close External Calibration VI) with the following
parameters to finish the calibration adjustment:
calHandle:
myCalHandle
action:
DAQmx_Val_Action_Commit
or
DAQmx_Val_Action_Cancel
Use the action
cancel
if there has been any error during the calibration or if you do not
want to save the new calibration constants in the device EEPROM.
Use the action
commit
if you want to save the new calibration constants in the device
EEPROM.
The device is now calibrated with respect to the external sources.