background image

PXIe-4480/4481 Getting Started Guide

|

© National Instruments

|

5

1.

Ensure the AC power source is connected to the chassis before installing the module. The 
AC power cord grounds the chassis and protects it from electrical damage while you install 
the module.

2.

Make sure the chassis is powered off.

Caution

To protect yourself, the chassis, and the module from electrical hazards, 

leave the chassis powered off until you finish installing the module.

3.

Inspect the slot pins on the chassis backplane for any bends or damage prior to installation. 
Do not install a module if the backplane is damaged.

4.

Remove the protective plastic covers from all the captive screws on the module front panel.

5.

Identify a supported slot in the chassis. The PXIe-4480/4481 module can be placed only in 
the supported slot types identified in Table 1. Figure 1 shows the slot type symbols of the 
various slot types. Refer to your chassis documentation for more details

Figure 1.  

PXI/PXIe Chassis Slot Type Symbol Location

6.

Touch any metal part on the chassis to discharge any static electricity.

7.

Ensure that the injector/ejector handle is in the unlatched (downward) position.

8.

Place the module edges into the module guides at the top and bottom of the chassis. Slide 
the module into the slot until it is fully inserted.

1

Slot Type Symbols Locations

Table 1.  

PXIe-4480/4481 Supported Slot Types

PXI Chassis Slot Type Symbol

Supported Slot Type Description

PXI Express Peripheral

PXI/PXI Express Hybrid Peripheral

PXI Express System Timing or Peripheral

NI PXIe-1062Q

1

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for PXIe-4480

Page 1: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Page 2: ...s product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility EMC stated in the product specifications These requirements and limits provide reasonabl...

Page 3: ...conductor to the chassis or earth ground at the load end of the cable For differential measurements where such a connection is inherently not possible the BNC cable can alternatively be wrapped in a s...

Page 4: ...ons Verifying the Kit Contents The following items are necessary to set up and use the PXIe 4480 4481 PXIe 4480 4481 analog input module NI DAQmx installation media PXIe 4480 4481 Getting Started Guid...

Page 5: ...acks for your operating system 3 Insert a compatible version of NI DAQmx To download the NI DAQmx driver visit ni com info and enter the Info Code daqmx 4 Follow the instructions in the installation p...

Page 6: ...he chassis The PXIe 4480 4481 module can be placed only in the supported slot types identified in Table 1 Figure 1 shows the slot type symbols of the various slot types Refer to your chassis documenta...

Page 7: ...l to the chassis using the front panel mounting screws Note Tightening the top and bottom mounting screws increases mechanical stability and also electrically connects the front panel to the chassis w...

Page 8: ...all modules installed in the chassis Your default module names may vary Note If you do not see your hardware listed press F5 to refresh the list of installed modules If the module is still not listed...

Page 9: ...S6 S7 S8 S9 S10 S11 S12 S13 S14 S15 S16 S17 S18 S19 S20 S21 S22 S23 S24 S25 S26 S27 S28 S29 S30 S31 S32 S33 S34 S35 S36 S37 S38 S39 S40 S41 S42 S43 S44 S45 S46 S47 S48 EX 0 EX 2 EX 3 EX 4 EX 5 EX 1 E...

Page 10: ...able for signal names and descriptions and the correlation to the pin numbers of the BNC plug Table 2 RJ50 Connector Pin Assignments RJ50 Modular Plug Pin Number Diagram Pin Signal Name Signal Descrip...

Page 11: ...l as other environmental information not included in this document Waste Electrical and Electronic Equipment WEEE EU Customers At the end of the product life cycle all products must be sent to a WEEE...

Page 12: ...nufacturer s declaration of conformity This system affords the user protection for electromagnetic compatibility EMC and product safety You can obtain the DoC for your product by visiting ni com certi...

Page 13: ...ments EULAs and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the NI global trade compliance policy...

Page 14: ...service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE IN...

Reviews: