40
|
ni.com
|
NI 5114 Calibration Procedure
You have finished adjusting the NI 5114. Repeat the steps listed in the
section to
reverify the performance of the NI 5114 after adjustments.
Table 11.
NI 5114 Input Parameters for External Adjustment
Iteration
Range (V)
Input (V)
1
40
18
2
20
9
3
10
4.5
4
4
1.8
5
2
0.9
6
1
0.45
7
0.4
0.18
8
0.2
0.09
9
0.1
0.045
10
0.04
0.018
11
40
-18
12
20
-9
13
10
-4.5
14
4
-1.8
15
2
-0.9
16
1
-0.45
17
0.4
-0.18
18
0.2
-0.09
19
0.1
-0.045
20
0.04
-0.018