Appendix A
Specifications
©
National Instruments Corporation
A-3
NI PXIe-8101/8102 User Manual
Environment
Maximum altitude .................................. 2,000 m (800 mbar)
(at 25 °C ambient temperature)
Pollution Degree .................................... 2
Indoor use only.
Caution
Clean the NI PXIe-8101/8102 with a soft nonmetallic brush. Make sure that the
device is completely dry and free from contaminants before returning it to service.
Operating Environment
NI PXIe-8101/8102
Ambient temperature range ............ 5 to 50 °C
1
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3
high temperature limit.)
Relative humidity range.................. 10% to 90%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
Caution
The operating temperature must not be exceeded, even when used in a chassis
with a higher temperature range.
Storage Environment
NI PXIe-8101/8102
Ambient temperature range ............ –40 to 65 °C
(Tested in accordance with
IEC-60068-2-1 and
IEC-60068-2-2. Meets
MIL-PRF-28800F Class 3
low temperature limit.)
Relative humidity range.................. 5% to 95%, noncondensing
(Tested in accordance with
IEC-60068-2-56.)
1
Processor should not throttle CPU frequency under reasonable, worst case processor work loads in high operating
temperatures.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com