National Instruments Corporation
3
NI PXI-8109 Installation Guide
Figure 2 shows an NI PXI-8109 controller installed in the system controller
slot of a National Instruments PXI-1042Q chassis. You can place
PXI devices in any other slot.
Figure 2.
NI PXI-8109 Installed in a PXI Chassis
How to Remove the Controller from the PXI Chassis
The NI PXI-8109 controllers are designed for easy handling. To remove
the unit from the PXI chassis:
1.
Turn off power.
2.
Remove cables from the controller front panel.
3.
Loosen the bracket-retaining screws in the front panel.
4.
Press the injector/ejector handle down.
5.
Slide the unit out of the chassis.
What If the NI PXI-8109 Does Not Boot?
Several problems can cause a controller not to boot. Here are some things
to look for and possible solutions.
Things to Notice
•
Which LEDs come on? The
PWR OK/FAULT
LED should stay lit
green. The
Drive
LED should blink during boot as the disk is accessed.
•
What appears on the display? Does it hang at some particular point
(BIOS, Operating System, etc.)? If nothing appears on the screen,
try a different monitor. Does your monitor work with a different PC?
If it hangs, note the last screen output that you saw for reference when
consulting National Instruments technical support.