background image

5.

Power on the chassis.

Note

 

If you are using a MXI controller, power on the chassis before powering

on the computer.

6.

Verify that the module appears in MAX.

7.

Reset the module in MAX and perform a self-test.

If the ACCESS LED still fails to light and failures continue, contact NI technical support or
visit 

ni.com/support

.

Related Information

Installing the PXI-5153

 on page 5

Installing the PCI-5153

 on page 6

What Should I Do if the PXI/PCI-5153 Doesn't Appear
in MAX?

1.

In the MAX configuration tree, expand 

Devices and Interfaces

.

2.

Expand the 

Chassis

 tree to see the list of installed hardware, and press <F5> to refresh

the list.

3.

If the module is still not listed, power off the system, ensure that all hardware is correctly
installed, and restart the system.

4.

Navigate to the Device Manager.

Operating System

Description

 

 

Windows 10/8.1

Right-click the Start button, and select 

Device Manager

.

Windows 7

Select 

Start

»

Control Panel

»

Device Manager

.

5.

If you are using a PXI or PXI Express controller, verify that a 

National Instruments

entry appears in the 

System Devices

 list. Reinstall NI-SCOPE and the module if error

conditions appear in the list. If you are using an MXI controller, right-click 

PCI-to-PCI

Bridge

, and select 

Properties

 from the shortcut menu to verify that the bridge is enabled.

What Should I Do if the Module Fails the Self-Test?

1.

Restart the system.

2.

Launch MAX, and perform the self-test again.

3.

Power off the chassis.

4.

Reinstall the failed module in a different slot.

5.

Power on the chassis.

6.

Perform the self-test again.

Thermal Shutdown Error

If you receive an over-temperature or thermal shutdown error and your device shuts down,
complete the following steps to re-enable your device:

14

  | 

ni.com

  | 

PXI/PCI-5153 Getting Started Guide

Summary of Contents for PCI-5153

Page 1: ...manner not described in this document Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Kit Contents 3 Other Equipment 3 Preparing the Enviro...

Page 2: ...ion Furthermore any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules Verifying the Syste...

Page 3: ...l items not included in your kit to install or operate your device PXI Devices A PXI chassis a PXI SCXI combination chassis or a PXI CompactPCI chassis with a controller and the chassis documentation...

Page 4: ...068 2 1 and IEC 60068 2 2 Relative humidity range 10 to 90 noncondensing Tested in accordance with IEC 60068 2 56 Maximum altitude 2 000 m at 25 C ambient temperature Pollution Degree 2 Indoor use onl...

Page 5: ...slot pins on the chassis backplane for any bends or damage prior to installation Do not install a module if the backplane is damaged 4 Remove the black plastic covers from all the captive screws on th...

Page 6: ...imize cooling air flow depending on your application 12 Power on the chassis Note For more information about maximizing cooling air flow refer to the Maintain Forced Air Cooling Note to Users included...

Page 7: ...anufacturers use a securing lever made of plastic to secure PCI devices such a lever is unacceptable and must be removed Use the screw provided in the kit to screw down the device Otherwise you must u...

Page 8: ...CH 0 CH 1 TRIG PFI 1 PFI 0 ACCESS ACTIVE CH0 CH1 TRIG PFI0 PFI1 Table 1 PXI PCI 5153 Front Panel Signal Connections Connector Description Function CH 0 CH 1 Standard BNC female connector Analog input...

Page 9: ...PE 1 Launch MAX 2 In the configuration tree expand Devices and Interfaces to see the list of installed NI hardware Installed modules appear under the name of their associated chassis 3 Expand your Cha...

Page 10: ...libration capabilities which you can access programmatically with NI SCOPE and your ADE or interactively with NI SCOPE SFP or MAX Related Information Making a Measurement on page 12 What Should I Do i...

Page 11: ...Microsoft Visual C C Use examples located in the file directory Users Public Documents National Instruments NI SCOPE examples You can modify an NI SCOPE C example to create an application with Micros...

Page 12: ...he acquisition display 5 If the SFP is not already running click Run Making a Measurement with LabVIEW 1 Launch LabVIEW 2 Select Help Find Examples 3 Open the example VI that you want to use by select...

Page 13: ...to NI RTSI Cable and select the device you want to add For more information about connecting PCI devices refer to NI High Speed Digitizers Help Programming Reference NI TClk Synchronization Help Troub...

Page 14: ...e Device Manager Operating System Description Windows 10 8 1 Right click the Start button and select Device Manager Windows 7 Select Start Control Panel Device Manager 5 If you are using a PXI or PXI...

Page 15: ...Where to Go Next NI PXIe 4112 EXPLORE LEARN CREATE DISCOVER Located online at ni com manuals the application development environment ADE for your application about hardware features or review device...

Page 16: ...hone support outside the United States visit the Worldwide Offices section of ni com niglobal to access the branch office websites which provide up to date contact information support phone numbers em...

Reviews: