NI USB-9219 User Guide and Specifications
10
ni.com
Wiring the NI USB-9219 Device
Caution
A high voltage screw terminal backshell must be installed when using hazardous
voltages (>42.4 V
pk
, 60 VDC).
Note
Table 2 illustrates the accessories available from
ni.com
for use with the
NI USB-9219.
The NI USB-9219 has four 6-terminal connectors that provide connections
for four analog input channels. Connect the positive signal of the signal
source to the positive input signal terminal (HI) and the negative signal of
the signal source to the negative input signal terminal (LO). Use the
excitation terminals if your sensor requires a separate excitation
connection. Refer to Table 3 for the signal names and Table 4 for the
terminal assignments for each mode. Refer to the
section for information about connections in each mode.
Table 2.
Accessories
Accessory
Accessory Description
NI 9972
Strain relief for 6-position spring terminal
NI 9973
6-position connectors
Table 3.
Signal Names
Module
Terminal
Signal Name
Signal Description
1
T+
TEDS Data
2
T–
TEDS COM
3
EX+/HI
*
Positive excitation or input
signal
4
HI
Positive input signal
5
EX–/LO
*
Negative excitation or input
signal
6
LO
Negative input signal
*
Depending on the mode, terminals 3 and 5 are either the excitation signals or the input signals.
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
1
2
3
4
5
6
Ch 0
Ch 1
Ch 2
Ch 3