Figure 4. Connecting Non-Referenced Single-Ended Signals to the NI USB-7855R
PGIA
ADC
AI+
NI USB-7855R
Connection
Accessory
AISENSE
AIGND
V
cm
V
1
+
–
+
–
Overvoltage
Protection
AI–
Overvoltage
Protection
Overvoltage
Protection
Connecting Differential Voltage Signals
You can connect grounded or floating differential signal sources to the NI USB-7855R.
Connect the positive voltage signal to the AI+ and the negative voltage signal to AI-. To
connect grounded differential signals to the NI USB-7855R, you must also connect the signal
reference to AI GND.
Figure 5. Connecting Grounded Differential Signals to the NI USB-7855R
PGIA
ADC
AI+
NI USB-7855R
Connection
Accessory
AISENSE
AIGND
V
cm
V
1
+
–
+
–
Overvoltage
Protection
AI–
Overvoltage
Protection
Overvoltage
Protection
To connect floating differential signals to the NI USB-7855R, you must connect the negative
and positive signals to AI GND through 1 MΩ resistors to keep the voltage within the
common-mode voltage range. If the voltage source is outside the common-mode voltage
range, the NI USB-7855R does not read data accurately.
NI USB-7855R User Manual | © National Instruments | 5