Table 3. Autonomous Mode Capabilities
Capability
Description
Use the Open/Close FPGA
Reference VI functions to run
an FPGA VI and then
disconnect the USB cable.
Call the Close FPGA VI Reference Function without
aborting or resetting the FPGA VI before you disconnect
the USB cable.
1
After you disconnect the USB cable, any VI running on the
NI USB-7855R continues to run and collect data, which
can later be retrieved by reconnecting the USB cable and
re-opening the original FPGA reference.
Use Interactive Front Panel
Communication to run an
FPGA VI and then
disconnecting the USB cable.
The front panel indicators stop updating.
After you disconnect the USB cable, any VI running on the
NI USB-7855R continues to run. Restore the Interactive
Front Panel Communication by reconnecting the USB
cable.
Download the bitfile to flash
memory and set it to run when
loaded to FPGA.
Download the bitfile to flash memory and set it to run when
loaded to FPGA. The bitfile automatically starts running
whenever power is applied to the NI USB-7855R. To
collect data, you must reconnect the NI USB-7855R to a
host computer before powering off the device.
Refer to the
LabVIEW FPGA Module Help
for more
information about downloading an FPGA VI to the flash
memory.
Where to Go for Support
The National Instruments Web site is your complete resource for technical support. At
ni.com/
support
you have access to everything from troubleshooting and application development self-
help resources to email and phone assistance from NI Application Engineers.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world to
help address your support needs. For telephone support in the United States, create your
service request at
ni.com/support
and follow the calling instructions or dial 512 795 8248. For
telephone support outside the United States, visit the Worldwide Offices section of
ni.com/
niglobal
to access the branch office Web sites, which provide up-to-date contact information,
support phone numbers, email addresses, and current events.
1
The Host VI errors out if you do not call the Close FPGA VI Reference Function without aborting
or resetting the FPGA VI.
Refer to the NI Trademarks and Logo Guidelines at
ni.com/trademarks
for information on National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents
covering National Instruments products/technology, refer to the appropriate location:
Help
»
Patents in your software, the
patents.txt
file on your media, or the National Instruments Patent Notice at
ni.com/patents
. You can find information about
end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export
Compliance Information at
ni.com/legal/export-compliance
for the National Instruments global trade compliance policy and
how to obtain relevant HTS codes, ECCNs, and other import/export data.
© 2013 National Instruments. All rights reserved.
375942A-01
Nov13