NI SCXI-1339 Installation Instructions

Long-nose pliers

Wire cutter

Wire insulation stripper

3. Connect Signals

To connect the signal(s) to the terminal block, refer to Figures 1 and 2 
while completing the following steps:

Caution

This module is rated for Measurement Category I and intended to carry signal 

voltages no greater than 150 V. This module can withstand up to 800 V impulse voltage. 
Do not use this module for connection to signals or for measurements within Categories II, 
III, or IV. Do not connect to MAINS supply circuits (for example, wall outlets) of 115 or 
230 VAC. Refer to the 

NI Switches Getting Started Guide

 for more information on 

measurement categories.

When hazardous voltages (>42.4 V

pk

/60 VDC) are present on any relay terminal, safety 

low-voltage (

42.4 V

pk

/60 VDC) cannot be connected to any other relay terminal.

1.

Prepare the signal wire by stripping the insulation no more than 7 mm 
from the end of the wire.

2.

Remove the top cover screw.

3.

Unsnap and remove the top cover.

4.

Loosen the two strain-relief screws on the strain-relief bar.

5.

Run the signal wires through the strain-relief opening.

6.

Insert the stripped end of the wire fully into the terminal. Secure the 
wire by tightening the screw of the terminal. No bare wire should 
extend past the screw terminal. Exposed wire increases the risk of a 
short-circuit causing a failure.

7.

Connect the safety earth ground to the safety ground lug.

8.

Tighten the two screws on the strain-relief assembly to secure the 
cables.

9.

Reinstall the top cover.

10. Replace the top cover screw.

Summary of Contents for NI SCXI -1339

Page 1: ...ntions The following conventions are used in this guide The symbol leads you through nested menu items and dialog box options to a final action The sequence File Page Setup Options directs you to pull...

Page 2: ...To avoid damage in handling the terminal block take the following precautions Caution Never touch the exposed pins of connectors Ground yourself using a grounding strap or by touching a grounded obje...

Page 3: ...rement categories When hazardous voltages 42 4 Vpk 60 VDC are present on any relay terminal safety low voltage 42 4 Vpk 60 VDC cannot be connected to any other relay terminal 1 Prepare the signal wire...

Page 4: ...NI SCXI 1339 Installation Instructions 4 ni com Figure 1 SCXI 1339 Top Cover Diagram 1 Top Cover 2 Top Cover Screw 1 2...

Page 5: ...e following steps Note Install the SCXI 1129 if you have not already done so Refer to the NI Switches Getting Started Guide for more information 1 Plug the SCXI 1339 onto the front connector of the SC...

Page 6: ...1339 Terminal Block Specifications Maximum Working Voltage Maximum working voltage refers to the signal voltage plus the common mode voltage Channel to earth 150 V Installation Category I Channel to c...

Page 7: ...safety for electrical equipment for measurement control and laboratory use IEC 61010 1 EN 61010 1 UL 3111 1 UL 61010B 1 CAN CSA C22 2 No 1010 1 Note For UL and other safety certifications refer to th...

Page 8: ...he patents txt file on your CD or ni com patents 2001 2007 National Instruments Corporation All rights reserved 372792C Nov07 CE Compliance This product meets the essential requirements of applicable...

Reviews: