NI SCXI-1336 Installation Instructions

5. Expand the Number of Rows

The SCXI-1336 offers convenient methods for expanding the number of 
rows of a matrix using two or more SCXI-1336 terminal blocks.

Caution

When using matrix expansion plugs, only connect terminal blocks of the same 

type. Connecting different types of terminal blocks can damage the module or the 
equipment connected to the module. For example, you can connect two or more 
SCXI-1336 terminal blocks; however, do 

not

 connect the SCXI-1336 with any other type 

of terminal block.

Figure 4 shows how to use a matrix expansion plug with SCXI terminal 
blocks to expand the number of rows of a 4

×

32 matrix, as an example. 

The matrix expansion plug connects columns to expand the number of 
rows. To expand the number of rows of a matrix, connect a matrix 
expansion plug to the top or bottom column connectors of the two terminal 
blocks.

Figure 4.  

Expanding the Number of Rows with the Matrix Expansion Plug

1

Matrix Expansion Plug

2

Thumbscrews

3

Column Connectors

4

32

4

32

8

32

3

1

2

3

Module 1

Matrix

Expansion

Plug

Module 2

Summary of Contents for NI SCXI-1336

Page 1: ...ermine when to install the terminal block Visit ni com switches for information on other switching solutions Conventions The following conventions are used in this guide The symbol leads you through n...

Page 2: ...riables filenames and extensions and code excerpts 1 Unpack the Terminal Block To avoid damage in handling the terminal block take the following precautions Caution Never touch the exposed pins of con...

Page 3: ...e for more information on measurement categories When hazardous voltages 42 4 Vpk 60 VDC are present on any relay terminal safety low voltage 42 4 Vpk 60 VDC cannot be connected to any other relay ter...

Page 4: ...NI SCXI 1336 Installation Instructions 4 ni com Figure 1 SCXI 1336 Top Cover Diagram 1 Top Cover 2 Top Cover Screw 1 2...

Page 5: ...I 1336 Installation Instructions Figure 2 SCXI 1336 Parts Locator Diagram 1 Column Connectors 2 Screw Terminals 3 Rear Connector 4 Thumbscrew 5 Strain Relief Screw 6 Strain Relief Bar 7 Safety Ground...

Page 6: ...the SCXI 1129 if you have not already done so Refer to the NI Switches Getting Started Guide for more information 1 Plug the SCXI 1336 onto the front connector of the SCXI 1129 2 Tighten the top and...

Page 7: ...ect two or more SCXI 1336 terminal blocks however do not connect the SCXI 1336 with any other type of terminal block Figure 4 shows how to use a matrix expansion plug with SCXI terminal blocks to expa...

Page 8: ...2 ADC 2 AAC Environmental Operating temperature 0 to 50 C Storage temperature 20 to 70 C Humidity 10 to 90 RH noncondensing Pollution Degree 2 Approved at altitudes up to 2 000 m Indoor use only Safet...

Page 9: ...elded cabling CE Compliance This product meets the essential requirements of applicable European Directives as amended for CE marking as follows Low Voltage Directive safety 73 23 EEC Electromagnetic...

Page 10: ...al Instruments trademarks Other product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products refer to the a...

Reviews: