background image

Table 4.

  NI 5646R-G Front Panel LEDs

LED

Indications

 

 

ACCESS Indicates the basic hardware status of the device.

Off—The device is not yet functional or has detected a problem with a PXI
Express power rail.

Amber—The device is being accessed. 

Accessed

 means that you are writing to

the device setup registers to control the device, reading from the device to
monitor the device status, or transferring data to/from the device.

Green—The device is controllable through the software.

ACTIVE When using NI-RFSG, the ACTIVE LED indicates the state of the device.

Off—The device is idle.

Solid green—The device is generating a waveform.

Solid red—The device has detected an error. The LED remains red until the error
condition is removed.

Related Information

Refer to your device specifications document for more information about front panel
connectors and LEDs.

Configuring the NI 5646R-G in MAX

Use Measurement & Automation Explorer (MAX) to configure your National Instruments
hardware. MAX informs other programs about which devices reside in the system and how
they are configured. MAX is automatically installed with NI-RFSG.
1.

Launch MAX.

2.

In the configuration tree, double-click 

Devices and Interfaces

 to see the list of installed

devices.

Installed devices appear under the name of their associated chassis.

3.

Expand your 

Chassis

 tree item.

MAX lists all devices installed in the chassis. Your default device names may vary.

Note

 

If you do not see your device listed, press <F5> to refresh the list of

installed devices. If the device is still not listed, power off the system, ensure
the device is correctly installed, and restart.

4.

Record the device identifier MAX assigns to the hardware. Use this identifier when
programming the NI 5646R-G.

NI PXIe-5646R-G Getting Started Guide

  | 

© National Instruments

  | 

9

Summary of Contents for NI PXIe-5646R-G

Page 1: ...mpaired if it is used in a manner not described in this document Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Verifying the Kit Contents...

Page 2: ...Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance operate this product only with cables le...

Page 3: ...ument Other Equipment There are several required items not included in your device kit that you need to install or operate the NI 5646R G Required Items A PXI Express chassis and chassis documentation...

Page 4: ...ator to install NI software on your computer 1 Install an ADE such as LabVIEW or LabWindows CVI 2 Insert the driver software media into your computer The installer should open automatically If the ins...

Page 5: ...chassis to rise above the optimal operating temperature for the device potentially causing thermal shutdown shorter lifespans or improper performance 5 Position the chassis so that inlet and outlet ve...

Page 6: ...ting screws increases mechanical stability and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 13 Cover all empty slots u...

Page 7: ...r switch routed to the NI 5646R G ensure no signal transients are sourced to the NI 5646R G Hardware Front Panel Connectors and Indicators Figure 4 NI 5646R G Front Panel DIGITAL I O LVTTL REF OUT 1 6...

Page 8: ...erence Clock REF OUT Output terminal that can export a 10 MHz Reference Clock or the 250 MHz Sample Clock PFI 0 Programmable function digital I O DIO connector for use with triggers or events DIGITAL...

Page 9: ...more information about front panel connectors and LEDs Configuring the NI 5646R G in MAX Use Measurement Automation Explorer MAX to configure your National Instruments hardware MAX informs other prog...

Page 10: ...temperature has drifted more than 5 C from the temperature at the last self calibration To periodically adjust for small performance drifts that occur with product aging NI recommends you perform the...

Page 11: ...cation in Microsoft Visual C C Related Information For detailed instructions about how to generate signals in a specific ADE refer to the Getting Started section of the NI RF Signal Generators Help Re...

Page 12: ...ED on the NI 5646R G hardware module illuminates 5 Click RF On Off to stop signal generation Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical su...

Page 13: ...sing a PXI controller verify that a National Instruments entry appears in the system device list Reinstall NI RFSG and the device if error conditions appear in the list If you are using an MXI control...

Page 14: ...software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located using the NI Example Finder Located at ni...

Page 15: ...your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressw...

Page 16: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Reviews: