You must install the software before installing the hardware.

Before you install the hardware, refer to the guidelines in the 

Maintain Forced-Air Cooling

Note to Users

 included with the module to ensure that the device can cool itself effectively.

This document is also available at 

ni.com/manuals

.

The NI 5646R-G is a three-slot module with two backplane connectors. The module must be
installed into three adjacent chassis slots, and the left two slots must be PXI Express
compatible.
1.

Ensure the AC power source is connected to the chassis before installing the module.

The AC power cord grounds the chassis and protects it from electrical damage while you
install the module.

2.

Power off the chassis.

3.

Inspect the slot pins on the chassis backplane for any bends or damage prior to
installation. Do not install a module if the backplane is damaged.

4.

If the chassis has multiple fan speed settings, ensure the fans are set to the highest setting.

Note

 

Inadequate air circulation could cause the temperature inside the chassis

to rise above the optimal operating temperature for the device, potentially
causing thermal shutdown, shorter lifespans, or improper performance.

5.

Position the chassis so that inlet and outlet vents are not obstructed.

Caution

 

Do not disconnect the cable that connects CAL IN to CAL OUT.

Removing the cable from or tampering with the CAL IN or CAL OUT front
panel connectors voids the product calibration and specifications are no longer
warranted.

6.

Remove the black plastic connectors from all the captive screws on the module front
panel.

7.

Identify a supported slot in the chassis. The following figure shows the symbols that
indicate the slot types.

Figure 2.

 Chassis Compatibility Symbols

NI PXIe-1062Q

1

2

3

4

5

1. PXI Express System Controller Slot
2. PXI Peripheral Slot
3. PXI Express Hybrid Peripheral Slot

4. PXI Express System Timing Slot
5. PXI Express Peripheral Slot

The NI 5646R-G can be placed in PXI Express peripheral slots, PXI Express Hybrid
peripheral slots, or PXI Express system timing slots.

8.

Touch any metal part of the chassis to discharge static electricity.

9.

Ensure that the ejector handle is in the unlatched (downward) position.

NI PXIe-5646R-G Getting Started Guide

  | 

© National Instruments

  | 

5

Summary of Contents for NI PXIe-5646R-G

Page 1: ...mpaired if it is used in a manner not described in this document Contents Electromagnetic Compatibility Guidelines 2 Verifying the System Requirements 2 Unpacking the Kit 2 Verifying the Kit Contents...

Page 2: ...Caution To ensure the specified EMC performance operate this product only with shielded cables and accessories Caution To ensure the specified EMC performance operate this product only with cables le...

Page 3: ...ument Other Equipment There are several required items not included in your device kit that you need to install or operate the NI 5646R G Required Items A PXI Express chassis and chassis documentation...

Page 4: ...ator to install NI software on your computer 1 Install an ADE such as LabVIEW or LabWindows CVI 2 Insert the driver software media into your computer The installer should open automatically If the ins...

Page 5: ...chassis to rise above the optimal operating temperature for the device potentially causing thermal shutdown shorter lifespans or improper performance 5 Position the chassis so that inlet and outlet ve...

Page 6: ...ting screws increases mechanical stability and also electrically connects the front panel to the chassis which can improve the signal quality and electromagnetic performance 13 Cover all empty slots u...

Page 7: ...r switch routed to the NI 5646R G ensure no signal transients are sourced to the NI 5646R G Hardware Front Panel Connectors and Indicators Figure 4 NI 5646R G Front Panel DIGITAL I O LVTTL REF OUT 1 6...

Page 8: ...erence Clock REF OUT Output terminal that can export a 10 MHz Reference Clock or the 250 MHz Sample Clock PFI 0 Programmable function digital I O DIO connector for use with triggers or events DIGITAL...

Page 9: ...more information about front panel connectors and LEDs Configuring the NI 5646R G in MAX Use Measurement Automation Explorer MAX to configure your National Instruments hardware MAX informs other prog...

Page 10: ...temperature has drifted more than 5 C from the temperature at the last self calibration To periodically adjust for small performance drifts that occur with product aging NI recommends you perform the...

Page 11: ...cation in Microsoft Visual C C Related Information For detailed instructions about how to generate signals in a specific ADE refer to the Getting Started section of the NI RF Signal Generators Help Re...

Page 12: ...ED on the NI 5646R G hardware module illuminates 5 Click RF On Off to stop signal generation Troubleshooting If an issue persists after you complete a troubleshooting procedure contact NI technical su...

Page 13: ...sing a PXI controller verify that a National Instruments entry appears in the system device list Reinstall NI RFSG and the device if error conditions appear in the list If you are using an MXI control...

Page 14: ...software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located using the NI Example Finder Located at ni...

Page 15: ...your product supports calibration you can obtain the calibration certificate for your product at ni com calibration National Instruments corporate headquarters is located at 11500 North Mopac Expressw...

Page 16: ...and third party legal notices in the readme file for your NI product Refer to the Export Compliance Information at ni com legal export compliance for the National Instruments global trade compliance p...

Reviews: