DMM Settings
Calibrator Settings
As-Found Limits (V)
Range
Coupling
Amplitude
Frequency
Lower
Upper
5 V
DC
5 V
30 Hz
4.99450
5.00550
AC / DC
5 V
50 Hz
4.99650
5.00350
5 V
1 kHz
4.99650
5.00350
50 V
5 V
1 kHz
4.9875
5.0125
5 V
5 V
20 kHz
4.99650
5.00350
5 V
50 kHz
4.99450
5.00550
5 V
100 kHz
4.97400
5.02600
5 V
300 kHz
4.84750
5.15250
50 V
DC
50 V
30 Hz
49.9450
50.0550
AC /DC
50 V
50 Hz
49.9650
50.0350
50 V
1 kHz
49.9650
50.0350
300 V
50 V
1 kHz
49.915
50.085
50 V
50 V
20 kHz
49.9650
50.0350
50 V
50 kHz
49.9450
50.0550
50 V
100 kHz
49.7400
50.2600
50 V
300 kHz
48.4750
51.5250
300 V
DC
219 V
30 Hz
218.751
219.249
AC / DC
300 V
50 Hz
299.790
300.210
219 V
1 kHz
218.831
219.170
219 V
20 kHz
218.831
219.170
219 V
50 kHz
218.743
219.257
219 V
100 kHz
217.845
220.155
70 V
300 kHz
67.750
72.250
You have completed verifying the AC voltage mode of the PXIe-4080/4082. Select one of the
following options:
PXIe-4080/4082 Calibration Procedure
|
© National Instruments
|
9