Chapter 2

Hardware Overview

NI PXI-562x User Manual

2-4

ni.com

Incorporating the DDC

Optionally, you can route the data through the DDC before storing it in
onboard memory.

The DDC is a digital signal processing (DSP) chip, the Intersil
HSP50214B. The first stage uses a digital quadrature mixer that shifts a
signal to baseband from any frequency within the range of the digitizer.
The next stage decimates (reduces the sample rate) by an integer from
4–16,384. A series of programmable digital lowpass filters prior to each
stage of decimation prevents aliasing when the sample rate is reduced. You
can retrieve the decimated data as in-phase and quadrature, or as phase and
magnitude. A discriminator allows you to take the derivative of the phase
to demodulate an FM signal.

By mixing, filtering, and decimating the sampled data, the DDC allows you
to zoom in on a band of frequencies much narrower than the Nyquist band
of the ADC. The lower sample rate means that signals of longer duration
can be stored in the same amount of memory. For spectral analysis, you can
use a smaller, faster FFT to look at only the band passed through the DDC.

Refer to the

NI-SCOPE VI Reference Help

for specific DDC attributes you

can use to program your NI 562

x

. For more information on using the

onboard DDC with LabVIEW, refer to the online help included with
NI-SCOPE and the Spectral Measurements Toolset software.

Storing Data in Memory

Samples are acquired into onboard memory on the NI 562

x

before being

transferred to the host computer. The minimum size for a buffer is
approximately 256 samples although you can specify smaller buffers in
software. When specifying a smaller buffer size, the minimum number
of points are still acquired into onboard memory, but only the specified
number of points are retrieved into the host computer memory.

During the acquisition, samples are stored in a circular buffer that is
continually rewritten until a trigger is received. After the trigger is received,
the NI 562

x

continues to acquire posttrigger samples if you have specified

a posttrigger sample count. The acquired samples are placed into onboard
memory. The number of posttrigger or pretrigger samples is limited only by
the amount of onboard memory.

Summary of Contents for NI PXI-562 Series

Page 1: ...Computer Based Instruments NI PXI 562x User Manual High Speed Frequency Domain Digitizer NI PXI 562x User Manual July 2002 Edition Part Number 322949C 01...

Page 2: ...14 24 24 Germany 089 741 31 30 Greece 01 42 96 427 Hong Kong 2645 3186 India 91 80 4190000 Israel 03 6393737 Italy 02 413091 Japan 03 5472 2970 Korea 02 3451 3400 Malaysia 603 9596711 Mexico 001 800...

Page 3: ...ents installation operation or maintenance instructions owner s modification of the product owner s abuse misuse or negligent acts and power failure or surges fire flood accident actions of third part...

Page 4: ...ments could void the user s authority to operate the equipment under the FCC Rules Class A Federal Communications Commission This equipment has been tested and found to comply with the limits for a Cl...

Page 5: ...ought for OEMs if also available from an original manufacturer that also markets in the EU or where compliance is not required as for electrically benign apparatus or cables To obtain the DoC for this...

Page 6: ...3 Safety Information 1 4 Chapter 2 Hardware Overview How the NI 562x Works 2 1 Connecting Signals 2 2 Conditioning the Signal Impedance Dither Gain and AC Coupling 2 3 Input Impedance 2 3 Dither 2 3 D...

Page 7: ...on The sequence File Page Setup Options directs you to pull down the File menu select the Page Setup item and select Options from the last dialog box This icon denotes a note which alerts you to impor...

Page 8: ...x ni com Related Documentation The following documents contain information that you might find helpful as you read this manual NI PXI 5620 Specifications NI PXI 5621 Specifications NI SCOPE User Manua...

Page 9: ...e NI PXI 5621 Specifications documents for NI PXI 562x performance specifications This chapter provides information on installing connecting signals to and acquiring data from the NI 562x modules The...

Page 10: ...installation does not start automatically navigate to your CD drive and click setup exe Caution You must turn off and unplug your chassis before installing your device To prevent damage due to electr...

Page 11: ...sfully installed and configured the necessary software and hardware to use your NI 562x Acquiring Data Programmatically You can acquire data programmatically either by writing an application for your...

Page 12: ...an explosive atmosphere or where there may be flammable gases or fumes Operate the product only at or below the pollution degree stated in the NI PXI 5620 Specifications and the NI PXI 5620 Specifica...

Page 13: ...stallation Category II are measurements on household appliances portable tools and similar equipment Installation Category III is for measurements performed in the building installation This category...

Page 14: ...Chapter 1 Taking Measurements with the NI PXI 562x NI PXI 562x User Manual 1 6 ni com Below is a diagram of a sample installation...

Page 15: ...r are applied to the signal Refer to the Conditioning the Signal Impedance Dither Gain and AC Coupling section for more information 3 The ADC converts the signal from analog to digital Refer to the Di...

Page 16: ...connectors and an SMB connector One of the SMA connectors INPUT is for attaching the analog input signal you want to measure The second SMA connector REF CLK IN is a 50 10 MHz AC coupled reference inp...

Page 17: ...Rin is the input impedance of the NI 562x Rs is the output impedance of the external device If the signal you are measuring has an output impedance other than 50 your measurements are affected by thi...

Page 18: ...amount of memory For spectral analysis you can use a smaller faster FFT to look at only the band passed through the DDC Refer to the NI SCOPE VI Reference Help for specific DDC attributes you can use...

Page 19: ...performs frequency translation filtering and decimation after signals go through the ADC Refer to the Incorporating the DDC section for more information The PLL uses a phase detector to synchronize t...

Page 20: ...you can configure the NI 562x to begin another acquisition that is stored in another memory record on the device This process is a multiple record acquisition To perform multiple record acquisitions...

Page 21: ...ignals with a minimum pulse width requirement of 100 ns or 16 ns times the DDC decimation Figure 2 5 Digital Trigger Sources Calibration Although the NI 562x is factory calibrated it needs periodic ca...

Page 22: ...he following timing and triggering features that you can use for synchronizing multiple digitizers System Reference Clock A 10 MHz clock on the PXI backplane with 100 ppm accuracy It is independently...

Page 23: ...Assisted Support Options Contact NI engineers and other measurement and automation professionals by visiting ni com ask Our online system helps you define your question and connects you to the experts...

Page 24: ...ossary Prefix Meanings Value p pico 10 12 n nano 10 9 micro 10 6 m milli 10 3 k kilo 103 M mega 106 G giga 109 Symbols percent positive of or plus negative of or minus per degree plus or minus ohm les...

Page 25: ...of signal conditioning that improves accuracy in the resulting digitized signal and reduces noise amplitude flatness a measure of how close to constant the gain of a circuit remains over a range of f...

Page 26: ...ng a default parameter value recorded in the driver In many cases the default input of a control is a certain value often 0 that means use the current default setting differential input an analog inpu...

Page 27: ...are trying to measure G gain the factor by which a signal is amplified sometimes expressed in decibels H hardware the physical components of a computer system such as the circuit boards plug in boards...

Page 28: ...level the relative priority at which a device can interrupt ISA industry standard architecture L LabVIEW Laboratory Virtual Instrument Engineering Workbench a program development application based on...

Page 29: ...e overrange a segment of the input range of an instrument outside of the normal measuring range Measurements can still be made usually with a degradation in specifications P PCI Peripheral Component I...

Page 30: ...in proportions or in percent of full scale For example a system has 12 bit resolution one part in 4 096 resolution and 0 0244 of full scale rms root mean square a measure of signal amplitude the squa...

Page 31: ...according to temperature See also thermal drift thermal drift measurements that change as the temperature varies thermal EMFs thermal electromotive forces voltages generated at the junctions of dissim...

Page 32: ...ally used with a PC that has the functionality of a classic stand alone instrument 2 a LabVIEW software module VI which consists of a front panel user interface and a block diagram program Vrms volts...

Page 33: ...1 D data acquisition multi record acquisitions 1 3 2 6 programmatically 1 3 data storing in memory 2 4 DDC digital downconverter incorporating 2 4 overview 2 4 dead time in multiple record acquisitio...

Page 34: ...e 2 2 installing software and hardware 1 1 safety information 1 4 NI SCOPE driver 1 1 O online technical support A 1 P phase detector 2 5 phase locked loop PLL 2 5 phone technical support A 1 professi...

Page 35: ...pport A 1 TIO timing engine 2 6 training customer A 1 trigger and clock routing area 2 6 triggering digital trigger sources figure 2 7 overview 2 7 troubleshooting resources A 1 V voltage controlled c...

Reviews: