NI 5105 Calibration Procedure
Password
The default calibration password is
NI
.
Calibration Interval
External Calibration
The external calibration interval for the NI 5105 is two years; however, the measurement
accuracy requirements of your application determine how often you should perform external
calibration. You can shorten this interval based on your application. Refer to
for more information.
Self-Calibration
Self-calibration can be performed whenever necessary to compensate for environmental
changes.
Caution
Although you can use self-calibration repeatedly, self-calibrating the
NI 5105 more than a few times a day may cause excessive wear on the relays over
time.
NI-SCOPE LabVIEW Reference Help
Contains: LabVIEW programming concepts and reference information about
NI-SCOPE VIs and functions
NI-SCOPE Function Reference Help
Contains: Reference information for NI-SCOPE C functions and NI-SCOPE
C properties