NI PXI-4070 Calibration Procedure
54
ni.com
9.
Reset the calibrator for safety reasons.
10. Call
niDMM_CalAdjustMisc
with
Type
set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the AC voltage modes of the NI 4070. Select
one of the following options:
• If you want to perform additional adjustments, go to the
Adjusting
Current Modes
section.
•
If you do not want to perform the current mode adjustments, go to the
section to verify your new calibration
coefficients before saving them to the EEPROM.
•
If you do not want to verify the adjustments you have just made, go to
the
Completing the Adjustment Procedures
Adjusting Current Modes
If you do not use the current modes (DC and AC), or the accuracy is
insignificant for your application, you can skip this section and select one
of the following options:
•
If you skip this section and you want to verify the new calibration
coefficients before saving them to the EEPROM, repeat the
section (except for
•
If you skip this section and you do not want to verify the new
calibration coefficients, go to the
section.
39
50 V
300
NIDMM_VAL_AC_VOLTS
300
300e3
50 V
300
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
300e3
40
10 V
500
NIDMM_VAL_AC_VOLTS
300
500e3
10 V
500
NIDMM_VAL_AC_VOLTS_
DCCOUPLED
300
500e3
Table 10.
niDMM_CalAdjustACFilter Parameters (Continued)
Iteration
Calibrator Output
niDMM_CalAdjustACFilter Parameters
Amplitude
Frequency
(kHz)
Mode
Range
(V)
Frequency
(Hz)