NI PXI-4070 Calibration Procedure
46
ni.com
56. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
57. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
58. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
59. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
60. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
61. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
62. Call
niDMM_CalAdjustMisc
with
Type
set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the DC voltage and resistance modes of the
NI 4070. Select one of the following options:
•
If you want to perform additional adjustments, go to the
Voltage (AC- and DC-Coupled) Modes
section, or to the
section if you have opted not to adjust AC voltage.
•
If you are not performing the AC voltage or the current adjustments,
go to the
calibration coefficients before saving them to the EEPROM.
•
If you do not want to verify the adjustments you have just made, go to
the