background image

Safety Voltages

Connect only voltages that are within the following limits:

Maximum voltage

[1]

Channel-to-COM

±30 V maximum on one channel at a time,

Measurement Category I

Isolation voltages
Channel-to-channel

None

Channel-to-earth ground
Continuous

60 VDC, Measurement Category I

Withstand

1,000 Vrms, verified by a 5 s dielectric withstand

test

Measurement Category I is for measurements performed on circuits not directly

connected to the electrical distribution system referred to as MAINS voltage. MAINS

is a hazardous live electrical supply system that powers equipment. This category is

for measurements of voltages from specially protected secondary circuits. Such

voltage measurements include signal levels, special equipment, limited-energy

parts of equipment, circuits powered by regulated low-voltage sources, and

electronics.

Caution

 Do not connect the NI 9401 to signals or use for measurements

within Measurement Categories II, III, or IV.

Note

 Measurement Categories CAT I and CAT O are equivalent. These test

and measurement circuits are not intended for direct connection to the

MAINS building installations of Measurement Categories CAT II, CAT III, or

CAT IV.

1

 The maximum voltage that can be applied or output between any channel and

COM without damaging the module or other devices.

Safety Guidelines for Hazardous Locations

ni.com

4

NI-9401

Summary of Contents for NI 9401

Page 1: ...NI 9401 2023 04 25...

Page 2: ...tromagnetic Compatibility Guidelines 6 Preparing the Environment 7 Connecting the NI 9401 8 NI 9401 Signals 8 Ports 8 Connecting a Serial Peripheral Interface Device 8 Connecting Digital Devices 9 Ove...

Page 3: ...entire system 2005 2015 National Instruments Corporation All rights reserved Refer to the National Instruments _Legal Information directory for information about NI copyright patents trademarks warra...

Page 4: ...is category is for measurements of voltages from specially protected secondary circuits Such voltage measurements include signal levels special equipment limited energy parts of equipment circuits pow...

Page 5: ...for Class I Division 2 Caution For Division 2 and Zone 2 applications install the system in an enclosure rated to at least IP54 as defined by IEC EN 60079 15 Caution For Division 2 and Zone 2 applica...

Page 6: ...tection against harmful interference when the product is operated in the intended operational electromagnetic environment This product is intended for use in industrial locations However harmful inter...

Page 7: ...s Register certification for a product visit ni com certification and search for the LR certificate or look for the Lloyd s Register mark on the product Caution In order to meet the EMC requirements f...

Page 8: ...6 NC DIO7 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 NI 9401 Signals Table 1 Signal Descriptions Signal Description COM Common reference connection to isolated ground DIO Digita...

Page 9: ...V DIO1 COM DIO7 NI 9401 LED Switch DIO0 TTL Signal DIO3 0 Configured for Input DIO7 4 Configured for Output Overcurrent Short Circuit Protection The overcurrent protection allows only a specified amou...

Page 10: ...hnical support and ensures that you receive important information updates from NI A Declaration of Conformity DoC is our claim of compliance with the Council of the European Communities using the manu...

Page 11: ...websites which provide up to date contact information support phone numbers email addresses and current events National Instruments 11 NI 9401 2023 National Instruments Corporation...

Reviews: