NI 9264 Operating Instructions and Specifications
14
ni.com
Protection
Overvoltage ................................ ±27 V at 25 °C
Short-circ
u
it................................ Indefinitely
Update time
Noise ................................................. 500
μ
V
rms
Crosstalk ........................................... 90 dB
Settling time (100 pF load, to 1 LSB)
20 V step..................................... 20
μ
s
1 V step....................................... 15
μ
s
0.1 V step.................................... 13
μ
s
Number of
Channels
Update Time for
cRIO-9151 R Series
Expansion Chassis
Update Time for All
Other Chassis
1
3.7
μ
s min
3.1
μ
s min
2
6.6
μ
s min
5.3
μ
s min
3
9.4
μ
s min
7.5
μ
s min
16
47
μ
s min
37
μ
s min