NI 9244
AI0
Phase A
Phase B
Phase C
AI2
AI1
The 9244 cannot measure the entire tolerance range or high crest factor signals on
690 Vrms systems in this configuration.
Connecting Single-Phase Measurement Configurations
NI recommends the following single-phase connection types.
3-Wire Measurement (Split Phase) Configuration
NI 9244
AI0
Neutral
AI1
2-Wire Measurement
NI 9244
AI0
Neutral
Connection Guidelines
■
You must use 2-wire ferrules to create a secure connection when connecting
more than one wire to a single terminal on the NI-9244.
ni.com
10
NI-9244 Getting Started