NI 9234 Calibration Procedure

|

© National Instruments

|

9

e.

Disable the function generator.

f.

Clear the task.

5.

Calculate the phase for each channel on the NI 9234. NI recommends using the Extract 
Single Tone Information VI to calculate phase.

6.

Compare the channel-to-channel phase difference to the limits in Table 10.

7.

Repeat steps through 6 for each test point in Table 10.

If the Phase Matching verification procedure determines that the NI 9234 is outside of the limits, 
refer to 

Worldwide Support and Services

 for assistance in returning the device to NI.

Common-Mode Rejection Ratio Verification

Complete the following procedure to determine the As-Found status of the NI 9234.

1.

Connect the NI 9234 to the function generator and short the positive pin and negative shell 
of the NI 9234 channel. 

2.

Connect the function generator to the chassis ground as shown in Figure 4.

Figure 4.  

CMRR Verification Connections to the NI 9234

3.

Configure the function generator according to Table 11.

Table 10.  

NI 9234 Verification Test Limits for Phase Matching

Generator Frequency (kHz)

Phase Matching Channel-to-Channel Limit

1

±0.085°

10

±0.49°

Table 11.  

Function Generator Configuration for CMRR Verification

Signal Type

Amplitude (V

pk

)

Frequency (kHz)

Sinewave

±0.5

1

F

u

nction

Gener

a

tor

NI 92

3

4

+

AI

x

HI

LO

Summary of Contents for NI 9234

Page 1: ...NI 9234...

Page 2: ...ion 6 Phase Matching Verification 8 Common Mode Rejection Ratio Verification 9 IEPE Current Verification 11 IEPE Compliance Voltage Verification 12 Adjustment 14 Accuracy Adjustment 14 EEPROM Update 1...

Page 3: ...ctions and Specifications NI 9234 specific information specifications and calibration interval NI DAQmx Readme Operating system and application software support in NI DAQmx LabVIEW Help LabVIEW progra...

Page 4: ...tion generator capable of sourcing 1 kHz and 10 kHz 4 5 Vpk AC sinusoidal signals with amplitude uncertainty of 10 or less Phase Matching Use a function generator capable of sourcing both 1 kHz and 10...

Page 5: ...minutes to ensure that the NI 9234 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9234 1 Install NI DAQmx 2 Make sure the NI cDA...

Page 6: ...utput is not grounded floating Notes The analog inputs on the NI 9234 are grounded so the calibrator must be not grounded floating to prevent ground loop measurement errors To float the output of the...

Page 7: ...calibrator to Standby mode STBY 2 Connect the NI 9234 to the calibrator Refer to Figure 2 for a connection diagram Figure 2 Gain Matching Verification Connections to the NI 9234 3 Verify that the cali...

Page 8: ...calibrator to Standby mode STBY 8 Disconnect the calibrator from the device 9 Perform the following calculation using the recorded V1 and V2 values 10 Repeat steps 1 through 9 for each channel 11 Com...

Page 9: ...a Create an AI voltage task on the NI 9234 according to Table 8 Note Wait 20 s for the AC coupling filter on the NI 9234 to settle b Configure the AI voltage task timing according to Table 9 c Enable...

Page 10: ...e device to NI Common Mode Rejection Ratio Verification Complete the following procedure to determine the As Found status of the NI 9234 1 Connect the NI 9234 to the function generator and short the p...

Page 11: ...sing the recorded AmplitudeAcq h Compare the calculation result to the minimum CMRR limit of 40 dB 6 Repeat steps 1 through h for each channel If the Common Mode Rejection Ratio verification procedure...

Page 12: ...14 b Configure the AI voltage task timing according to Table 15 c Start the task d Record the DMM measurement e Clear the task 4 Compare the DMM measurement to the minimum current limit of 2 0 mA 5 R...

Page 13: ...2 mA 3 Measure IEPE Source short circuit current a Create an AI voltage task on the NI 9234 according to Table 16 b Configure the AI voltage task timing according to Table 17 c Start the task d Recor...

Page 14: ...ording to Table 19 c Set the source current on the SMU to the recorded SMUSourceCurrent value d Start the task e Enable the SMU f Record the SMU measurement SMUVoltage g Disable the SMU h Clear the ta...

Page 15: ...the DAQmx adjustment function according to Table 20 g Set the calibrator to Standby mode STBY h Repeat steps d through g for each calibration voltage in the array i Disconnect the NI 9234 from the ca...

Page 16: ...tion Conditions The following calibration specifications are valid under the following conditions Ambient temperature 23 5 C NI 9234 installed in slot 8 of an NI cDAQ 9178 chassis Slots 1 through 7 of...

Page 17: ...site is your complete resource for technical support At ni com support you have access to everything from troubleshooting and application development self help resources to email and phone assistance...

Reviews: