NI 9214 Operating Instructions and Specifications
30
ni.com
Table 3.
Thermocouple Type T/E Measurement Accuracy (°C)
–10
0 °
C
0 °
C
1
00 °
C
3
00 °
C
5
00 °
C
7
00 °
C
9
00 °
C
H
igh-
Re
sol
u
ti
o
n
Typical,
23 ± 5 °C
0.54
0.37
0.33
0.33
0.37
0.43
0.49
Max,
–20 to 70 °C
1.76
1.17
0.89
1.00
1.25
1.58
1.94
Max,
–40 to 70 °C
1.76
1.17
1.04
1.17
1.42
1.74
2.11
H
igh-
Spe
ed
Typical,
23 ± 5 °C
1.25
0.88
0.77
0.69
0.69
0.78
0.90
Max,
–20 to 70 °C
2.59
1.77
1.38
1.41
1.60
1.96
2.37
Max,
–40 to 70 °C
2.59
1.77
1.53
1.53
1.77
2.13
2.55
NI_9214.fm Page 30 Monday, November 1, 2010 9:52 AM