With autozero channel off
Types J, K, T, E, N ............... <0.05 °C
Type B .................................. <0.20 °C
Types R, S ............................ <0.45 °C
Figures 5, 6, 7, 8, and 9 show the typical and maximum errors
1
for
each thermocouple type when used with the NI 9211E over the full
temperature range. The figures also display the maximum errors
for the thermocouple types with the NI 9211E at room
temperature, which is 15 to 35 °C. The figures account for gain
errors, offset errors, differential and integral nonlinearity,
quantization errors, noise errors, and isothermal errors. The figures
do not account for the accuracy of the thermocouple itself.
1
The maximum errors assume an isothermal error of 1.0 °C. Refer to the
Cold-Junction Temperature Measurement Accuracy
section for more information.