NI 9207 Calibration Procedure

|

© National Instruments

|

9

Reverification

Repeat the 

Verification

 section to determine the As-Left status of the device.

Note

If any test fails Reverification after performing an adjustment, verify that you 

have met the 

Test Conditions

 before returning your device to NI. Refer to 

Worldwide 

Support and Services

 for assistance in returning the device to NI.

Accuracy Under Calibration Conditions

The values in the following table are based on calibrated scaling coefficients, which are stored 
in the onboard EEPROM.

The following accuracy table is valid for calibration under the following conditions:

Ambient temperature 23 °C ± 5 °C

NI 9207 installed in slot 8 of an NI cDAQ-9178 chassis

Slots 1 through 7 of the NI cDAQ-9178 chassis are empty

Note

The test limits listed in Tables 4 and 7 are derived using the values in 

Table 10.

Note

For operational specifications, refer to the most recent 

NI 9207 Operating 

Instructions and Specifications

 online at 

ni.com/manuals

.

Table 10.  

NI 9207 Accuracy Under Calibration Conditions

Input Channels

Percentage of Reading

Percentage of Range

*

Voltage

0.100%

0.020%

Current

0.150%

0.012%

Range equals 10.4 V for voltage and 22 mA for current.

Summary of Contents for NI-9207

Page 1: ...4 Current Accuracy Verification 5 Adjustments 7 Voltage Accuracy Adjustment 7 Current Accuracy Adjustment 8 EEPROM Update 8 Reverification 9 Accuracy Under Calibration Conditions 9 Worldwide Support a...

Page 2: ...rating Instructions and Specifications NI 9207 specific information specifications and calibration interval NI DAQmx Help Information about creating applications that use the NI DAQmx driver LabVIEW H...

Page 3: ...me of at least 10 minutes to ensure that the NI 9207 measurement circuitry is at a stable operating temperature Initial Setup Complete the following steps to set up the NI 9207 1 Install NI DAQmx 2 Ma...

Page 4: ...ode OPR 5 Acquire and average samples a Create and configure an AI voltage task on the NI 9207 according to Table 2 b Configure the AI voltage task according toTable 3 Set the ADC timing mode to high...

Page 5: ...2 Connect the NI 9207 to the calibrator as shown in Figure 2 Figure 2 Current Accuracy Connections Note If the calibrator outputs are truly floating connect the negative output to a quiet earth ground...

Page 6: ...point in Table 7 10 Disconnect the calibrator from the device 11 Repeat steps 1 through 10 for each current channel on the NI 9207 Table 5 NI 9207 Configuration for Current Accuracy Verification Input...

Page 7: ...mmended calibration voltages for the NI 9207 d Set the calibrator to a reference value determined by the array of recommended calibration voltages e Set the calibrator to Operate mode OPR f Call and C...

Page 8: ...g to Table 9 g Set the calibrator to Standby mode STBY h Repeat steps d through g for each calibration current in the array i Disconnect the calibrator from the NI 9207 Ensure that there are no connec...

Page 9: ...s which are stored in the onboard EEPROM The following accuracy table is valid for calibration under the following conditions Ambient temperature 23 C 5 C NI 9207 installed in slot 8 of an NI cDAQ 917...

Page 10: ...private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52 227 14 DFAR 252 227 7014 and DFAR 252 227 7015 Worldwide Support and Services The Nat...

Reviews: