NI 6110/6111/6115/6120 Calibration Procedure

12

ni.com

gpctrNum

0

action

ND_RESET

3.

Call 

GPCTR_Set_Application

 to configure the counter for 

pulse-train generation. Set the following parameters:

deviceNumber

—The value obtained from MAX

gpctrNum

0

application

ND_PULSE_TRAIN_GNR

 

4.

Call 

GPCTR_Change_Parameter

 to configure the counter to output a 

pulse with an off time of 100 ns. Set the following parameters:

deviceNumber

—The value obtained from MAX

gpctrNum

0

paramID

ND_COUNT_1

paramValue

2

5.

Call 

GPCTR_Change_Parameter

 to configure the counter to output a 

pulse with an on time of 100 ns. Set the following parameters:

deviceNumber

—The value obtained from MAX

gpctrNum

0

paramID

ND_COUNT_2

 

paramValue

2

6.

Call 

Select_Signal

 to route the counter signal to the 

GPCTR0_OUT

 

pin on the device connector. Set the following parameters:

deviceNumber

—The value obtained from MAX

signal

ND_GPCTR0_OUTPUT

source

ND_GPCTR0_OUTPUT

sourceSpec

ND_LOW_TO_HIGH

7.

Call 

GPCTR_Control

 to start the generation of a square wave. Set the 

following parameters:

deviceNumber

—The value obtained from MAX

gpctrNum

0

action

ND_PROGRAM

8.

The device begins to generate a 5 MHz square wave when the 

GPCTR_Control

 function completes execution. Compare the value 

read by the counter to the test limits in the appropriate table. If the 
value falls between these limits, the device has passed the test.

9.

Disconnect the counter from the device.

You have now verified the counter on the device.

Summary of Contents for NI 6110

Page 1: ...PXI 6120...

Page 2: ...equipment software calibration functions and environment needed for calibration Calibration Process This section provides step by step instructions for verifying and calibrating your device NI 6110 6...

Page 3: ...often you should calibrate the device NI recommends performing a complete calibration at least once every year You can shorten this interval based on the demands of your application Equipment and Oth...

Page 4: ...o create applications that use NI DAQ These documents are your primary references for writing a calibration utility For more information on the products you are calibrating refer to the device user ma...

Page 5: ...connections and test conditions during calibration use the following guidelines Keep connections to the device as short as possible Long cables and wires can act as antennae picking up extra noise tha...

Page 6: ...re These flowcharts only refer to the appropriate NI DAQ function calls You may want to refer to the Calibration Process section to the NI DAQ Function Reference Help and to the NI DAQ User Manual for...

Page 7: ...able for the product determine the channel output polarity and voltage to verify AO_VWrite deviceNumber chan voltage GPCTR_Control deviceNumber chan ND_RESET GPCTR_Set_Application deviceNumber chan ND...

Page 8: ...rocedure describes how to configure a device with NI DAQ 1 Install NI DAQ The latest version is available at ni com 2 Power off the computer and install the device in an available slot 3 Power on the...

Page 9: ...on Note Calibrator connections depend on the channel you are verifying The verification process requires you to individually test each channel 2 Use the following guide to determine the connections be...

Page 10: ...l AI_Configure to inform NI DAQ of the input mode input range and input polarity selected for the device Set the following parameters deviceNumber The value obtained from MAX chan The channel currentl...

Page 11: ...values 11 Make the appropriate connections to verify the next channel by repeating steps 2 and 3 and proceed from step 4 You have completed verifying the analog input levels on the device Verifying A...

Page 12: ...and lower limits on the data sheet If the value falls between these limits the device has passed this test 7 Repeat steps 4 6 to test all values 8 Disconnect the DMM from DAC0OUT and reconnect it to D...

Page 13: ...umber The value obtained from MAX gpctrNum 0 paramID ND_COUNT_2 paramValue 2 6 Call Select_Signal to route the counter signal to the GPCTR0_OUT pin on the device connector Set the following parameters...

Page 14: ...input channel 0 pin 68 3 Connect the negative output of the calibrator to analog input channel 0 pin 34 and to ACH0GND pin 67 4 Call Get_Cal_Date to obtain the date of the last calibration The functio...

Page 15: ...output operation To do this repeat the procedure in the Verifying the Performance of the NI 6110 6111 6115 6120 section NI 6110 6111 6115 6120 Specifications The tables in this section contain the sp...

Page 16: ...or output to verify performance This value is divided into two columns Location and Value Location refers to where the test value fits within the test range Pos FS is positive full scale and Neg FS i...

Page 17: ...8951 9 9050 9 8950 5 Bipolar 2 PosFS 4 95000 4 94741 4 95259 4 94734 4 95266 5 Bipolar 2 Zero 0 0000 0 00083 0 00083 0 00083 0 00083 5 Bipolar 2 NegFS 4 95000 4 95259 4 94741 4 95266 4 94734 2 Bipolar...

Page 18: ...800 41 3995 41 7605 41 3987 41 7613 42 Bipolar 0 2 Zero 0 0000 0 0366 0 0366 0 0366 0 0366 42 Bipolar 0 2 NegFS 41 5800 41 7605 41 3995 41 7613 41 3987 20 Bipolar 0 5 PosFS 19 8000 19 7319 19 8681 19...

Page 19: ...ipolar Zero 0 000000 0 008900 0 008900 0 008900 0 008900 20 Bipolar NegFS 9 990000 10 003266 9 976734 10 003435 9 976565 Table 10 Counter Values for the NI 6115 Counter Values Set Point MHz Lower Limi...

Page 20: ...03 0 008303 0 008303 1 Bipolar 10 NegFS 0 990000 0 999859 0 980141 0 999876 0 980124 0 5 Bipolar 20 PosFS 0 4950000 0 485919 0 504081 0 485911 0 504089 0 5 Bipolar 20 Zero 0 000000 0 008303 0 008303 0...

Page 21: ...NI 6110 6111 6115 6120 Calibration Procedure 20 ni com Table 13 Counter Values for the NI 6120 Counter Values Set Point MHz Upper Limit MHz Lower Limit MHz 5 4 9995 5 0005...

Reviews: