NI 5922 Calibration Procedure
4
ni.com
You may find the following documentation helpful as you write your
calibration procedure:
•
NI High-Speed Digitizers Getting Started Guide
•
NI High-Speed Digitizers Help
•
NI PXI-5922 Specifications
•
NI-SCOPE Function Reference Help
•
NI-SCOPE VI Reference Help
You can download the latest versions of all documentation from the NI Web
site at
ni.com/manuals
.
Self-Calibration Procedures
The NI 5922 includes precise internal circuits and references used during
self-calibration to adjust for time and temperature drift. The digitizer gain,
offset, and flatness are all corrected in self-calibration. Measuring the
accuracy of these internal calibration circuits with another instrument
during external calibration provides traceability for the verification
procedure. Absolute accuracy is ensured by compensating for any error
measured in the internal references.
Note
Self-calibrate the digitizer before you perform verification. NI-SCOPE includes
self-calibration example programs for LabVIEW, CVI, and Microsoft Visual C.
LabVIEW
(
version)\instr.lib\niScope
Calibrate\niScopeCal.llb
LabVIEW VI library containing VIs for calling
the NI-SCOPE calibration API; access
calibration functions from the NI-SCOPE
calibration section of the LabVIEW function
palette
IVI\Drivers\niScope\niScopeCal.fp
CVI function panel file that includes external
calibration function prototypes and help on
using NI-SCOPE in the CVI environment
IVI\Include\niScopeCal.h
Calibration header file, which you must include
in any C program accessing calibration
functions. This file automatically includes
niScope.h
, which defines the rest of the
NI-SCOPE interface
IVI\Drivers\niScope\Examples
Directory of NI-SCOPE examples for CVI, C,
Visual C++, and Visual Basic
Table 1.
Calibration File Locations after Installing NI-SCOPE 2.8 or Later (Continued)
File Name and Location
Description