Programs

»

National Instruments MXI-3

»

MXI-3 Optimization

. MXI-4 and

MXI-Express optimization are performed automatically by the hardware.

4.

Perform a device self-test to verify installation by right-clicking the device and selecting

Self-Test

.

A dialog box appears and indicates whether the device passed the test. When the self-test finishes,
a message indicates either successful verification or that an error occurred.

Note

If the device does not pass the self-test, reboot your system and repeat

steps 1 through 4. If the device still does not pass the self-test, refer to

ni.com/support

.

5.

Configure the device settings by completing the following steps:

a.

Record the device number or device name assigned to the device. You need this number when
programming the device.

The assigned device name is appended to the device in its configuration tree label. For
example, after installing the NI 5421, the device configuration tree label may appear as

NI PXI-5421: <Dev1>

, where

Dev1

is the device name. When developing your

application, the resource name for your device is the device name MAX assigned to the
device. You can rename NI-DAQmx devices by right-clicking the device and selecting

Rename

.

b.

PCI only—Right-click the NI signal generator in the list and select

Properties

.

c.

PCI only—Select the appropriate tab to set the configuration options. The configuration options
you see depend on the device.

d.

Click

OK

to close the window.

Setting up Synchronization in MAX

If you plan to share triggers or clocks to synchronize the devices, you must identify or configure certain
components in MAX.

Note

The following sections contain information required for any type of synchronization,

including NI-TClk synchronization. For information about NI-TClk synchronization, refer to

Programming

»

NI-TClk Synchronization Help

in the

NI Signal Generators Help

.

PXI Devices

Identify the PXI system controller by completing the following steps:

1.

In the MAX Configuration tree under

Devices and Interfaces

, right-click

PXI System

»

Identify As

and select your controller from the list. For example, select

External PC

if you are using a MXI

controller in an external PC.

2.

Expand the PXI System tree and select the chassis name to identify the chassis you are using.

PXI Express devices controlled using MXI-Express

Identify the PXI Express system controller in the MAX Configuration tree under

Devices and Interfaces

by right-clicking

PXI System

»

Identify As

, and selecting

External PC

. Your chassis is automatically

identified.

PXI Express devices controlled using a PXI Express controller

PXI Express controllers installed in a PXI Express chassis are automatically identified in MAX.

NI Signal Generators Getting Started Guide

10

ni.com

Summary of Contents for NI 5402

Page 1: ...rent versions of documentation visit ni com manuals For the latest version of NI FGEN visit ni com idnet Contents Conventions 2 Related Documentation 3 Verifying the System Requirements 3 Unpacking 4...

Page 2: ...a warning advising you to take precautions to avoid electrical shock When symbol is marked on a product it denotes a component that may be hot Touching this component may result in bodily injury Bold...

Page 3: ...lp from within the SFP For device specific help select Help Hardware Specific Help in the SFP FGEN Soft Front Panel Help Contains information on configuring and testing data acquisition DAQ devices us...

Page 4: ...I FGEN DVD which includes the NI Signal Generators Help NI Analog Waveform Editor CD Eligible hardware kits NI 5412 5421 5422 5441 5442 5450 5451 only receive an activation license NI Modulation Toolk...

Page 5: ...as LabVIEW or LabWindows CVI 2 Insert the NI FGEN DVD The NI FGEN installer should open automatically Note If the installation window does not appear navigate to the DVD drive double click the drive...

Page 6: ...esetDevice function Perform a device reset in MAX The Measurement Automation Explorer Help for NI DAQmx contains information about resetting a device in MAX Review the guidelines in the Maintain Force...

Page 7: ...6 Uninstalling PXI and PXI Express Modules When removing modules you must power down the chassis Hot Surface During operation the metal surfaces of devices may become hot Be careful when removing the...

Page 8: ...ured devices may affect the accuracy of device specifications Some computer manufacturers use a securing lever made of plastic to secure PCI devices such a lever is unacceptable and must be removed Us...

Page 9: ...the accuracy of the signal generator specifications Configuring and Testing in MAX Complete the following steps to configure and test your device 1 Launch MAX 2 In the configuration pane double click...

Page 10: ...t Properties c PCI only Select the appropriate tab to set the configuration options The configuration options you see depend on the device d Click OK to close the window Setting up Synchronization in...

Page 11: ...complete the following steps Note The NI 5450 does not support standard waveform generation You can use arbitrary waveform generation when using the NI 5450 and the FGEN SFP 1 Launch the FGEN SFP fro...

Page 12: ...ownload a waveform to the device onboard memory 6 Select Load Waveform from disk The Select File dialog box opens and displays waveform example files You also can download a waveform by selecting File...

Page 13: ...ory where NIDocDir is one of the following directories Windows 2000 XP Documents and Settings All Users Shared Documents National Instruments Windows Vista Users Public Documents National Instruments...

Page 14: ...ion tables that describe the signal connection options for NI signal generators The NI Signal Generators Help includes signal descriptions connector pin assignments and routing information NI PXI PCI...

Page 15: ...pts a PLL Reference clock from an external source and can frequency lock the Sample clock timebase to the external Reference clock Input REF IN Provides a TTL level output of the waveform being genera...

Page 16: ...nectors as shown in the following figure Signal descriptions are listed in the following table SINE REF OUT REF IN PFI 0 CH 0 CLOCK ACCESS ACTIVE NI PXI 540 4 100 MHz Freq Source Figure 7 NI PXI 5404...

Page 17: ...ase to the external Reference clock Input REF IN NI PXI PCI 5412 5421 5422 5441 The NI 5412 is a 100 MS s 20 MHz 14 bit arbitrary waveform generator The NI 5421 is a 100 MS s 43 MHz 16 bit arbitrary w...

Page 18: ...lable on the NI 5412 or the NI 5421 5422 with the 8 MB memory option Table 5 NI PXI PCI 5412 5421 5422 5441 Front Panel Connectors Function Access Connector Provides the waveform output Output CH 0 Ac...

Page 19: ...om several clock event and trigger sources Input Output PFI 1 Routes the 16 bit digital pattern outputs digital pattern clock output trigger outputs trigger inputs and a clock input Input Output DIGIT...

Page 20: ...rence clock from an external source and can frequency lock the Sample clock timebase to the external Reference clock The signal on this connector can also be used as a Sample clock source Input CLK IN...

Page 21: ...waveform generator with onboard signal processing OSP The NI 5451 is a two slot PXI Express 400 MS s 135 MHz dual channel 16 bit arbitrary waveform generator with onboard signal processing OSP and 16...

Page 22: ...or can route signals from several clock event and trigger sources Input Output PFI 1 Where to Go for Support The National Instruments Web site is your complete resource for technical support At ni com...

Page 23: ...srael 972 0 3 6393737 Italy 39 02 413091 Japan 81 3 5472 2970 Korea 82 02 3451 3400 Lebanon 961 0 1 33 28 28 Malaysia 1800 887710 Mexico 01 800 010 0793 Netherlands 31 0 348 433 466 New Zealand 0800 5...

Page 24: ...her product and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the appropriate locati...

Reviews: