Chapter 1
Introduction—Document Scope
©
National Instruments Corporation
1-3
Software
This section describes the software and documentation you need to
calibrate your NI 54
XX
device.
Writing Your Calibration Procedure
The calibration process is described in Chapter 3,
, including
step-by-step instructions on calling the appropriate calibration functions.
Be aware that many of the functions listed in Chapter 3,
, use
constants defined in the
niArbCal.h
file. To use these variables, you must
include
niArbCal.h
in your code. See Table 1-3 for file locations.
Table 1-2.
Equipment Required for Calibrating NI 54
XX
Devices
Instrument
Specifications
Recommended Equipment
5 1/2 Digit Digital Multimeter
(DMM)
At least 1 kHz bandwidth
NI 4060 or HP 34401A
Digital Counter
Ability to make duty cycle
measurements, measure
1 MHz frequency, and have
frequency measurement
accuracy
≤
1 ppm (accounting
for clock stability and
measurement error due to
resolution)
NI 6608 or HP 53131A with
appropriate timebase option
BNC Cable
50
Ω
—
BNC Connector (for signal
sources in PXI form factor)
50
Ω
—
SMB Connector (for signal
sources in PCI form factor)
50
Ω
—
50
Ω
Terminator
—
—
Spectrum Analyzer
Bandwidth > 32 MHz
Distortion < –70 dBc
Spurs to 32 MHz
HP 8590 or better
Power Meter or
Thermoconverter
100 kHz to 16 MHz equal or
better than ±2% or ±0.2 dB
—