National Instruments Corporation
45
NI myDAQ User Guide and Specifications
Weight
NI myDAQ device part number
195509D-01L and earlier................ 175.0 g (6.1 oz)
NI myDAQ device part number
195509E-01L and later ................... 164.0 g (5.8 oz)
Note
NI myDAQ device part number (
P/N: 195509
x
-01L
) is located on the product label
on the bottom of the device.
Screw-terminal wiring............................ 16 to 26 AWG
Torque for screw terminals .................... 0.22–0.25 N · m (2.0–2.2 lb · in.)
Environmental
The NI myDAQ device is intended for indoor use only.
Operating temperature
(IEC 60068-2-1 and IEC 60068-2-2) ..... 0 to 45 °C
Storage temperature
(IEC 60068-2-1 and IEC 60068-2-2) ..... –20 to 70 °C
Operating humidity
(IEC 60068-2-56) ................................... 10 to 90% RH, noncondensing
Storage humidity
(IEC 60068-2-56) ................................... 10 to 90% RH, noncondensing
Maximum altitude .................................. 2,000 m (at 25 °C ambient
temperature)
Pollution Degree (IEC 60664) ............... 2
Safety
Safety Voltages
Measurement Category I is for measurements performed on circuits not
directly connected to the electrical distribution system referred to as
MAINS
voltage. MAINS is a hazardous live electrical supply system that
powers equipment. This category is for measurements of voltages from
specially protected secondary circuits. Such voltage measurements include
signal levels, special equipment, limited-energy parts of equipment,
circuits powered by regulated low-voltage sources, and electronics.
Caution
Do not use this module for connection to signals or for measurements within
Measurement Categories II, III, or IV.