Switch Setting
Description
S6 S5 S4
S1
S2
S3
MIO with disabled temperature sensor mode (default
configuration)—Move switches S1, S2, S3, S4, S5, and S6 to the
positions shown on the left. In this mode:
•
The temperature sensor is not used.
•
AI 0 and AI 8 are available on screw terminals.
•
The signal accessory power LED (DS1) lights when the
MIO device is powered on and connected to the SCB-100.
Refer to the following figure for a detailed diagram.
S6 S5 S4
S1
S2
S3
MIO with single-ended temperature sensor mode—Move
switches S1, S2, S3, S4, S5, and S6 to the positions shown on the
left. In this mode:
•
The temperature sensor can be read using AI 0 in referenced
single-ended (RSE) mode.
•
AI 8 is available on a screw terminal.
•
The signal accessory power LED (DS1) lights when the
MIO device is powered on and connected to the SCB-100.
Refer to the following figure for a detailed diagram.
S6 S5 S4
S1
S2
S3
MIO with differential temperature sensor mode—Move switches
S1, S2, S3, S4, S5, and S6 to the positions shown on the left. In
this mode:
•
The temperature sensor can be read using AI 0 and AI 8 in
differential mode.
•
The signal accessory power LED (DS1) lights when the
MIO device is powered on and connected to the SCB-100.
Refer to the following figure for a detailed diagram.
S6 S5 S4
S1
S2
S3
DIO/passthrough mode—You can use DIO/passthrough mode
with an MIO device.
Refer to the
Using the SCB-100 with DIO/TIO Devices
configuration information and a detailed diagram.
NI SCB-100 Installation Guide and Specifications
|