NI 4070/4072 Calibration Procedure
46
ni.com
63. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
64. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e6
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
65. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
66. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
10e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
67. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
1e3
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
68. Call
niDMM_CalAdjustOffset
with the following parameters:
•
Mode
=
NIDMM_VAL_2_WIRE_RES
•
Range
=
100
•
Input Resistance
=
NIDMM_VAL_RESISTANCE_NA
69. Call
niDMM_CalAdjustMisc
with
Type
set to
NIDMM_EXTCAL_MISCCAL_SECTION
.
You have completed adjusting the DC voltage and resistance modes of the
NI 4070/4072. Select one of the following options:
•
If you are performing additional adjustments, refer to the following
sections, as applicable:
–
Adjusting AC Voltage (AC- and DC-Coupled) Modes
–
–
Adjusting Capacitance and Inductance (NI 4072 Only)
Caution
For the NI 4072, adjusting the capacitance and inductance is required. Skipping
this step causes an incorrect adjustment of the device.