National Instruments Corporation
17
NI 4070/4072 Calibration Procedure
14. Repeat step 13 for each of the remaining iterations shown in Table 3.
15. Output 500 mV at 30 Hz on the calibrator.
16. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function
=
NIDMM_VAL_AC_VOLTS_DCCOUPLED
•
Range
=
0.5
•
Resolution
=
500e–9
17. Call
niDMM_Read
. Verify that this measurement falls between the
limits listed in Table 18.
18. Refer to Table 4 for the appropriate calibrator outputs and parameter
values as you complete the following steps:
a.
On the calibrator, output the value listed in the
Calibrator Output
column in Table 4 for the current iteration.
b.
Call
niDMM_ConfigureMeasurement
with
Mode
set to
NIDMM_VAL_AC_VOLTS
and the remaining parameters as shown
in Table 4 for the current iteration.
Table 3.
niDMM_ConfigureMeasurement Parameters
Iteration
Calibrator Output
niDMM_ConfigureMeasurement Parameters
Amplitude
Frequency
Function
Range
Resolution
1
50 mV
50 Hz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
50 Hz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9
2
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9
3
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS
0.5
500e–9
50 mV
1 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.5
500e–9
4
50 mV
20 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
20 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9
5
50 mV
50 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
50 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9
6
50 mV
100 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
100 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9
7
50 mV
300 kHz
NIDMM_VAL_AC_VOLTS
0.05
50e–9
50 mV
300 kHz
NIDMM_VAL_AC_VOLTS_DCCOUPLED
0.05
50e–9