Table 11. LED State/Device Status
LED Color and
State
Device Status
Off
Device is reserved by NI-DAQmx but there is no activity, the device
is not powered, or the device is not reserved by NI-DAQmx.
Yellow, solid
Device firmware is booting or updating.
Red, blinking
Device has reset to factory default or the firmware image is corrupted.
If you are not performing a factory default reset, as listed in
the FieldDAQ to Factory-Default Settings
corrupted firmware.
Green, solid
Device is reserved and an NI-DAQmx task is running.
Resetting the FieldDAQ to Factory-Default
Settings
To reset your FieldDAQ device to factory-default settings, complete the following steps.
1.
Verify the power source is turned off.
2.
Connect Ethernet port 0 and port 1 with an Ethernet cable.
Figure 30. Connecting Port 0 and Port 1
0
10/100/1000
SYNC
1
LINK/ACT
3.
Turn on the power source to the device. After about 15 seconds, the STATUS LED blinks
red indicating that the device has been restored to factory-default settings.
4.
Reconnect the Ethernet port 0 to the host.
5.
Turn off and then turn on the power source to the device.
The FieldDAQ device factory-set defaults are listed in the following table.
Table 12. Device Default Settings
Attribute
Value
Hostname
FD11601-<
serial number
>
IP
DHCP or Link Local
40
|
ni.com
|