Table 1.
Hazardous Location Retention Accessories (Continued)
Port
Required Accessory
Part Number
Mini DisplayPort
Mini DisplayPort Cable Retention Bracket
156866-01
SD Card Slot
SD Door Kit
783660-01
Special Conditions for Hazardous Locations Use in Europe and
Internationally
The cDAQ-9132 has been evaluated as Ex nA IIC T4 Gc equipment under DEMKO
Certificate No. 12 ATEX 1202658X and is IECEx UL 14.0089X certified. Each device is
marked II 3G and is suitable for use in Zone 2 hazardous locations, in ambient temperatures
of -20 °C ≤ Ta ≤ -55 °C.
Caution
You must make sure that transient disturbances do not exceed 140% of
the rated voltage.
Caution
The system shall only be used in an area of not more than Pollution
Degree 2, as defined in IEC 60664-1.
Caution
The system shall be mounted in an ATEX/IECEx-certified enclosure with
a minimum ingress protection rating of at least IP54 as defined in IEC/EN 60079-15.
Caution
The enclosure must have a door or cover accessible only by the use of a
tool.
Power Requirements
Caution
The protection provided by the NI cDAQ-9132 can be impaired if it is
used in a manner not described in the user documentation.
Note
Some C Series modules have additional power requirements. For more
information about C Series module power requirements, refer to the
C Series module(s) documentation.
Note
Sleep mode for C Series modules is not supported in the NI cDAQ-9132.
Voltage input range
9 to 30 V (measured at the NI cDAQ-9132
power connector)
Maximum power consumption
1
40 W
Recommended power supply
100 W, 24 VDC
1
Includes maximum 1 W module load per slot across rated temperature and product variations.
NI cDAQ-9132 Safety, Environmental, and Regulatory Information
|
© National Instruments
|
3