ni.com

|

BNC-2120 Installation Guide

Connecting Differential Analog Input Signals

Use the BNC-2120 BNC connectors on the front panel to connect AI <0..7> signals to your 
DAQ device. The BNC-2120 is 

only

 intended for differential analog input signals. The number 

of connectors you use depends on your DAQ device and application. Complete the following 
steps to measure a differential (DIFF) analog input signal.
1.

Connect the BNC cable to one of the AI <0..7> BNC connectors on the front panel.

1

2.

If you are using AI 0, move the BNC/Temp. Ref. switch to the BNC position.
If you are using AI 1, move the BNC/Thermocouple switch to the BNC position.
If you are using AI 3, move the RES/BNC switch to the BNC position.

3.

Configure your software to measure the channel differentially.

4.

Move the FS/GS switch to the applicable position, depending on whether you are 
measuring a floating source (FS) or ground-referenced source (GS) analog input signal. 
Refer to th

Measuring Floating Signals

 an

Measuring Ground-Referenced Signals

 

sections for more information about these signal sources.

Measuring Floating Signals

To measure floating signal sources, move the corresponding switch below the BNC connector 
to the FS position. In the floating source switch position, the DAQ device’s amplifier negative 
terminal connects to ground through a 4.99 k

Ω

 resistor, as shown in Figure 2. Refer to your 

DAQ device documentation for more information about measuring floating signals.

Figure 2.  

Measuring a Floating Signal Source

1   

When using the BNC-2120 with Connector 1 of two connector M/X Series devices, the AI <0..7> BNCs 

on the BNC-2120 map to the AI <16..23> channels on the device.

+

S

ign

a

l

S

o

u

rce

BNC-2120

DAQ Device

(Differenti

a

l Inp

u

t Mode)

AI GND

+

4.99 k

Ω

F

S

F

S

 Mode

0.1 

μ

F

Summary of Contents for BNC-2120

Page 1: ...ing I O TIO connection Two user defined BNC connectors A function generator with the following outputs Frequency adjustable TTL compatible square wave Frequency and amplitude adjustable sine wave or t...

Page 2: ...Series User Manual the M Series User Manual the S Series User Manual or the X Series User Manual BNC cables 28 16 AWG wire Wire strippers Flathead screwdriver Installing the BNC 2120 Caution When this...

Page 3: ...BNC Connector 24 Power Indicator LED E L R T T S N I NA M U N O I A S T N _ Floating Source FS _ Ground Ref Source GS AO ANALOG OUTPUTS ANALOG INPUTS DIGITAL I O FUNCTION GENERATOR Frequency Selectio...

Page 4: ...damage the BNC 2120 all devices connected to it and the host computer Overvoltage can also cause an electric shock hazard for the operator National Instruments is not liable for damage or injury resu...

Page 5: ...revised its terminal names so they are easier to understand and more consistent among NI hardware and software products The revised terminal names used in this document are usually similar to the nam...

Page 6: ...Move the FS GS switch to the applicable position depending on whether you are measuring a floating source FS or ground referenced source GS analog input signal Refer to the Measuring Floating Signals...

Page 7: ...the temperature reference and thermocouple connector To use the integrated circuit IC temperature reference on the BNC 2120 move the BNC Temp Ref switch above the AI 0 BNC to Temp Ref The integrated c...

Page 8: ...renced single ended RSE mode 3 Connect the resistor to the RES and RES screw terminals You can measure resistors with values ranging from 100 to 1 M 4 Measure VCC on AI 3 and the voltage drop across t...

Page 9: ...onnecting signals to the screw terminals use 28 16 AWG wire with the insulation stripped to 0 28 in Table 2 describes the timing I O BNC and terminal signals on the BNC 2120 Table 2 BNC 2120 Timing I...

Page 10: ...to the general purpose counter 1 PFI 5 P1 5 Programmable Function Interface channel 5 or Port 1 Digital Input Output channel 5 AO SAMP CLK AO Sample Clock Signal As an output this pin is the ao Sample...

Page 11: ...pin can be used to route signals directly to the RTSI bus As an output this pin emits the Ctr0InternalOutput signal PFI 13 P2 5 Programmable Function Interface channel 13 or Port 2 Digital Input Outp...

Page 12: ...with counter 1 connect the PULSES terminal to PFI 3 and connect the UP DN terminal to the P0 7 digital I O terminal the up down pin of counter 1 Refer to Table 2 for more information about PFI 3 and P...

Page 13: ...6 shows an example of how to use the USER 1 and USER 2 BNCs To access the PFI 8 P2 0 CTR 0 SOURCE signal from a BNC connect USER 1 on the screw terminal block to PFI 8 P2 0 with a 28 16 AWG wire Figu...

Page 14: ...each digital line as listed in Table 3 Specifications This section lists the specifications of the BNC 2120 These specifications are typical at 25 C unless otherwise specified Note Refer to your DAQ...

Page 15: ...tions 28 16 AWG wire LED state indicators 8 1 each for lines P0 0 7 Protection DC max V Powered off 5 5 V Powered on 10 5 V Drive Vol 0 6 V 8 mA 1 6 V 24 mA Voh 4 4 V 8 mA 4 V 13 mA Function Generator...

Page 16: ...terminals 2 Output signals PULSES 96 pulses revolution UP DN High for clockwise rotation low for counterclockwise rotation Pulse width 1 s Power Requirement 5 VDC 5 200 mA sourced from the multifuncti...

Page 17: ...is used in a manner not described in this document Electromagnetic Compatibility This product meets the requirements of the following EMC standards for electrical equipment for measurement control an...

Page 18: ...NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers For additional environmental information refer to the Minimize Our En...

Page 19: ...t and company names mentioned herein are trademarks or trade names of their respective companies For patents covering National Instruments products technology refer to the appropriate location Help Pa...

Reviews: